This high specification process calibrator is lightweight, portable, battery or mains powered calibration unit and can be used to source and measure thermocouples, PT100s, μV, mV and mA.
When used with thermocouples, the 1090 can be used to measure/simulate temperature (in both Celsius and Fahrenheit) plus mV characteristics of type J, K, T, R, S, B, N and E type thermocoples. It also includes the ability to carry out PT100 measurement and simulations (based on 0.3850 alpha probe standard) between a range of -200 to 700°C.
The final measurement/simulation mode on this device enables it to measure 0 to ±30mV and 0 to ±60mA and source from 0 to ±80mV and 0 to ±80mA.
This calibration device also comes with cold junction compensation, meaning it can be operated with ot without internal cold junction compensation. Other additional features include a 24V process loop drive mode, inching (incrementing/decrementing) and built-in memory.
Time Electronics 1090 Portable Process Calibrator Key Features
- Measures and simulates 8 thermocouple types - J, K, T, R, S, B, N and E
- Measures and simulates PT100-RTD
- Measures and sources μV, mV and mA
- Switch between Celsius and Fahrenheit modes
- Battery or mains powered
- 10 point memory recall function
- Process loops 4 to 20mA and 0 to 50mA
- 24V loop driven voltage
- 60 hour typical battery life between charger
- Three levels of inching increments
Full List of Products Included
- Time Electronics 1090 Portable Process Calibrator
- Robust Carry Case
- Instruction Manual
Time Electronics 1090 Technical Specifications
Thermocouple Type | Temperature Range (°C) | Accuracy (°C) |
---|---|---|
J | -200 to 580 | 0.7 |
K |
-200 to -150 |
2.5 |
T |
-200 to 0 |
1.5 |
R |
50 to 400 |
3.0 |
S |
-50 to 100 |
3.0 |
B |
110 to 1000 |
3.5 |
N | -100 to 890 | 0.6 |
E | -50 to 400 | 0.4 |
Resolution |
0.1°C or 0.1°F |
Temperature Simulation |
---|
Thermocouple Type | Temperature Range (°C) | Accuracy (°C) |
---|---|---|
J |
-210 to 150 |
0.15 |
K |
-270 to 190 |
0.5 |
T |
-200 to 150 |
0.4 |
R |
-50 to 800 |
0.8 |
S |
-50 to 850 |
0.9 |
B |
100 to 1200 |
2.0 |
N |
-270 to 260 |
0.5 |
E | -50 to 1000 | 0.3 |
Resolution |
0.1°C or 0.1°F |
Note: An additional correction representing the equivalent 1μV should be allowed for stray thermal emf effects (Temperature measure & simulation).
PT100 Simulation |
---|
14 set temperature points (°C) | Accuracy |
---|---|
-100, -50, -20, 0, 20, 50, 100, 200, 300, 400, 500, 600, 700, 800 | 0.1% of resistance value (typically 0.5 °C) |
PT100 Measure |
---|
Range | Accuracy | Resolution |
---|---|---|
200 to 700 °C, 2 wire. | 0.2% of resistance value (typically 0.7 °C) | 0.2°C (or 0.2°F) |
Millivolt Measure |
---|
Range | Resolution | Accuracy | Input Resistance |
---|---|---|---|
0 to +/- 30mV |
10μV |
0.05% of full scale ±1 digit |
100kΩ |
Millivolt Source |
---|
Range | Resolution | Accuracy | Output Resistance |
---|---|---|---|
0 to 8mV | 0.5μV | +/- 4μV | 10Ω |
8 to 80mV | 5μV | 0.02% of full scale | 10Ω |
Milliamp Measure |
---|
Range | Resolution | Accuracy | Input Resistance |
---|---|---|---|
0 to 60mA |
20μA |
0.05% of full scale ±1 digit |
0.5Ω |
Milliamp Source |
---|
Range | Resolution | Accuracy |
---|---|---|
0 to 8mA | 0.5μA | +/-10μA |
8 to 80mA |
5μA |
0.02% of full scale |
Further Features and Functions |
---|
Inching | Three levels of increment, 0.1, 1 or 10 for °C/°F, and 1, 10, or 100 μV/ μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. |
24V Process Loop Drive |
A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA |
Memory recall & step |
10 memory locations for non-volatile storage of values. Manual & AutoStep, rate adjustable from 1 to 10 sec/step |
General Specifications
Cold Junction Compensation | Accuracy 0.2°C. Resolution 0.1°C. |
Operating Temperature | -10 to 40°C (15 to 105°F) |
Connections | Industry standard 4mm screw terminals |
Power | A Ni-MH rechargeable battery pack gives approx. 60 hours continuous use. The mains recharger supplied gives full recharge in 11 hours. To conserve battery life a user inactivity power-down feature is included. |
Dimensions | H235 x W150 x D75mm |
Weight | 1.2kg |
Supplied With | Carry case, rechargeable battery pack (installed) with mains charger, and user manual |
Optional Extras | Calibration certificates traceable NPL and UKAS |
Country of Origin | UK |