The Time 1077 contains 3 three modes: an adjustable current load for the simulation of a transducer on a line, an adjustable power supply mode from 14V to 40V and a precision current sourcing mode from 14 to 40V drive capacity.
This transducer simulator is primarily used for the testing and simulation of milliamp transducer systems both inside laboratories and when out in the field. To accommodate for tests outside the lab the 1077 is encased in protective rubber and is designed to resist common stresses such as those found in process control applications. It's also portable enough to be used in just one hand and won't require a desktop for comfortable use and it runs on rechargeable batteries too.
The 1077 includes fitted safety terminals and is made for use with standard 4mm shrouded plugs, standard normal plugs, bare wires and spade terminals.
Current Source and Load | |
Output |
0 to 100mA in 3 ranges: |
Accuracy | ± 0.02% of setting + ± 0.02% of range + ± 0.02μA |
Output Stability | Less than 60ppm/°C. Less than 25ppm per hour at constant temperatur |
Output Noise | Less than 15ppm of full scale |
Drive Voltage | Adjustable from 14 to 40 volts (Source mode) |
Drive Power | 2.4 watts maximum |
Applied Voltage | 3 volts minimum to 40 volts maximum (Load mode) |
Voltage Limit Indicator | Provides indication of insufficient drive voltage |
24 Volt Line Simulation | |
24V Line Simulation | Adjustable 14V to 40V, 100mA current limit, Maximum output power 2.4 watts |
Display | A 3.5 digit (1999 max) LCD display indicating line current |
Measure Range | 0 to 20ma |
Resolution | 10μA |
Accuracy | 0.2% of reading + 1 count |
General Specs | |
Dimensions | H200 x W75 x D110mm (H215 x W100 x D120mm including protective cover) |
Weight | 1.1kg (1.5kg including protective cover) |
Power Supply | NiMH rechargeable batteries with external mains recharger (supplied). Standard mains voltage is 220–250V, 50/60Hz. 100–125V, 50/60Hz is available but must be specified on ordering |