The TestSafe Medical Forehead and Body Temperature Infrared Thermometer is a non-contact, IR thermometer compliant with a range of medical standards, please see the 'Technical Specifications' for further details. It can be purchased with or without calibration.
Purchasing the TestSafe Medical Forehead and Body Temperature Infrared Thermometer with a calibration means it will be checked to ensure that it is working correctly and within its accuracy specifications. The resultant calibration certificate provides proof of this check and reassures the user that the TestSafe Medical Forehead and Body Temperature Infrared Thermometer can be relied upon to take repeatable temperature readings within its specified accuracy range.
The TestSafe Medical Forehead and Body Temperature Infrared Thermometer is designed to measure forehead/body temperature and surface/ object temperature. It can be used to identify persons with EBT (Elevated Body Temperature). This is often a sign of fever which is a symptom of many illnesses including the recent COVID-19 coronavirus. Persons with suspected EBT may need to undergo further virus/infection-specific medical screening. The TestSafe Medical Forehead and Body Temperature Infrared Thermometer can be used to streamline medical testing and can be used alongside medical procedures to help prevent the potential spread of infection.
When set to Surface mode, the TestSafe Medical Forehead and Body Temperature Infrared Thermometer has a temperature range of 0°C to 100°C; however, when set to Forehead/ Body mode, the thermometer has a temperature range of 32°C to 43°C and is able to measure skin temperatures to the following accuracies:
- 35.5°C to 42°C: ±0.2°C
- 32.0°C to 35.4°C and 42.1°C to 43°C: ±0.3°C
Measurements are displayed to a resolution of 0.1°C on the TestSafe Medical Forehead and Body Temperature Infrared Thermometer’s backlit display. The user can choose to view temperatures in °C or °F.
When set to Body/ Forehead mode, the colour of the TestSafe Medical Forehead and Body Temperature Infrared Thermometer's backlit display will change depending on the measured temperature:
- 32.0°C to 37.4°C = green
- 37.5°C to 37.9°C = orange
- 38.0°C to 43.0°C = red
This allows the user to quickly evaluate whether the person being screened is likely to have a fever. Please note that this feature is disabled when the TestSafe Medical Forehead and Body Temperature Infrared Thermometer is set to Surface mode. The individual being screened should stand between 3cm and 5cm away from the IR thermometer.
Additionally, the TestSafe Medical Forehead and Body Temperature Infrared Thermometer includes a user-settable Temperature Alarm function with an audible alert and a configurable Temperature Difference function.
Furthermore, temperatures are automatically saved to the TestSafe Medical Forehead and Body Temperature Infrared Thermometer's memory.
The TestSafe Medical Forehead and Body Temperature Infrared Thermometer provides a safe, easy and efficient method of ascertaining an individual’s body and forehead temperature.
TestSafe Medical Forehead and Body Temperature Infrared Thermometer Key Features
- A non-contact, IR thermometer compliant with a range of medical standards
- Measure forehead/ body and surface/ object temperature
- Identify persons with EBT, a sign of fever, which is a symptom of many illnesses
- Streamline medical testing
- Use alongside medical procedures to help prevent the potential spread of infection
- Surface mode temperature range: 0°C to 100°C
- Body/ Forehead mode temperature range: 32°C to 43°C
- Body/ Forehead mode accuracy between 35.5°C and 42°C: ±0.2°C
- Body/ Forehead mode accuracy between 32.0°C to 35.4°C and 42.1°C to 43°C: ±0.3°C
- Resolution: 0.1°C
- Selectable °C or °F measurement units
- 3-colour backlit display changes depending on the measured temperature (in Forehead/ Body mode)
- 32.0°C to 37.4°C = green backlight
- 37.5°C to 37.9°C = orange backlight
- 38.0°C to 43.0°C = red backlight
- User-settable Temperature Alarm function with an audible alert
- Configurable Temperature Difference function
- Temperatures are automatically saved to the data memory
- Persons being screened should stand between 3cm and 5cm from the thermometer
- Provides a safe, easy and efficient method of ascertaining an individual’s body/ forehead temperature
What’s Included?
- TestSafe Medical Forehead and Body Temperature Infrared Thermometer
TestSafe Medical Forehead and Body Temperature Infrared Thermometer Technical Specifications
Measurement mode | Forehead and object temperature modes |
Measuring range | For forehead temperature: 32-43ºC(89.6ºF-109.4ºF) |
For object surface temperature: 0-100ºC | |
Measuring accuracy: (At laboratory conditions) |
35.5°C to 42°C: ±0.2°C |
32.0°C to 35.4°C and 42.1°C to 43°C: ±0.3°C |
|
Clinical repeatability |
Within ±0.3°C |
Distance Meter | 3-5cm |
Display resolution | 0.1ºC/ 0.1°F |
Usage environment | 10-40ºC(50-104ºF),humidity 15-85%, 70-106KPa |
Storage environment | -25-50ºC(32-122ºF),humidity≤95% |
Power supply | 2 AAA Batteries |
Backlight | Highlight backlight |
Display | C/F |
Auto close | 15s |
Size | 149 (L) x 95mm (W) x 45mm |
Weight (g) | Approx. 145g (including battery) |
High Body Temperature hint | ≥38℃ |
Grade of waterproof | IP22 |
Electric shock | Internally powered ME equipment |
Applied part | Type BF applied part, including the whole unit |
Mode of operation | Continuous operation |
Product life | 5 years |
Software version | V0.1 |
Power supply | d.c.3V, 2 pcs AAA battery |
Note: | Not intended to be sterilized. |
Not for use in an OXYGEN RICH ENVIRONMENT |
TestSafe Medical Forehead and Body Temperature Infrared Thermometer Standard List
EN 980 | Symbols for use in the labeling of medical devices |
EN 1041 | Information supplied by the manufacturer with medical devices |
EN 60601-1 | Medical electrical equipment Part 1: General requirements for basic safety and essential performance |
EN 0601-1-6 | Medical electrical equipment ‒ Part1-6: General requirements for basic safety and essential performance ‒ Collateral standard: Usability |
EN 60601-1-11 | Medical electrical equipment ‒ Part 1-11: General requirements for basic safety and essential performance ‒ Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment |
EN 12470-5 | Clinical thermometers ‒ Part 5: Performance of infra-red ear thermometers (with maximum device) |
ISO 80601-2-56 | Medical electrical equipment part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement |
EN 62304 | Medical device software - Software life-cycle processes |
EN 62366 | Medical devices ‒ Application of usability engineering to medical devices |
EN ISO 10993-1 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process |
Reviews
SKU | TS-TM501 |
---|---|
Manufacturer Part Code | PC868 |
Country of Origin | China |