TP Solid / Dry Block-Temperature Calibrator
With the temperature calibrators of the TP Solid series, the main focus is on flexibility: In addition to dry block calibrators, they also include calibration baths, with which almost any temperature sensor can be calibrated irrespective of its shape. Both can be operated easily and intuitively. When being used as a fluid bath calibrator, the temperature sensors are directly immersed in the calibration liquid. This creates a direct temperature link between the calibrator and the devices under test without an insulating air gap. All TP Solid temperature calibrators are additionally equipped with a serial interface for computer-assisted monitoring of the calibration process. This flexibility in combination with the easy operation makes the TP Solid series ideal for use in machinery and plant engineering.
- ✓ Compact design for on-site use
- ✓ Large calibration volume for the simultaneous calibration of several DUT
- ✓ Ideal for temperature sensors with unusual shapes and dimensions
Whether for on-site use in the workshop, in the measuring and test room, or laboratory, SIKA temperature calibrators can be used everywhere. A wide range of industries and various application areas are covered.
- • Assembly, commissioning
- • Manufacturing, production
- • Maintenance, service
- • Quality assurance, test equipment monitoring
- • Repair
Calibration Bath Function
If temperature sensors with unusual shapes and dimensions are to be checked, the use of calibration liquids offers advantages. The DUT is placed directly into the liquid without insulating air gaps, thus creating a direct temperature seal between the calibrator and the DUT. The liquid, e.g. silicone oil, is selected according to the required calibration temperature. The continuously variable control of the magnetic stirrer, together with the pull-out sensor basket, causes vortices in the calibration liquid and thus creates a large homogenous temperature zone. Furthermore, the sensor basket ensures an unobstructed stirring operation and serves to protect the tank bottom.
Please contact us to customise your apparatus.
Linearisation Example: EPLIKTII100 | ||||
---|---|---|---|---|
With Linearisation | EPLIK | |||
Function | ||||
Direct filling | LII | |||
Medium | ||||
Water (2...95 ℃) | 01 | |||
Silicone 10 cSt (-35...155 ℃) | 10 | |||
Silicone 50 cSt (50...270 ℃) | 50 | |||
Linearisation Points | ||||
Standard | 0 | |||
Custom Specific | K |
Datasheets
Manuals & Guides
Certifications & Declarations
TP M165S-STD | TP M255S-STD-110V | TP M255S-STD | |
---|---|---|---|
Temperature Range | -35...165 °C at ambient temperature 20 °C | Room temperature...255 °C | |
Calibration Volume | Ø 60 x 170 mm | ||
Calibration Bath |
|||
Display Accuracy | ±0.1 °C | ±0.2 °C | |
Temperature Stability | ±0.05 °C | ||
Resolution of the Temperature Display | 0.01 °C in the range -9.99...99.99 °C, else 0.1 °C | ||
Reference Temperature Sensor | Internal, fixed installation | ||
Interface | RS485 (calibrator) to USB (PC) | ||
Connectivity | MODBUS | ||
Dimensions |
|||
→ Width | 210 mm | 150 mm | |
→ Height | 380 + 50 mm | 330 + 70 mm | |
→ Depth | 300 mm | 270 mm | |
Weight | Approx. 12.5 kg | Approx. 7.5 kg | |
Power Supply | 100...240 VAC, 50 / 60 Hz | 110...115 V 60 Hz | 230 V 50 Hz |
Power Consumption | Approx. 375 W | Approx. 1000 W | |
Display |
|||
Display | 2-line, 4-digit digital display red / green, unit °C / °F | ||
Approvals |
|||
Reviews
SKU | TP M |
---|