TP Solid / Dry Block-Temperature Calibrator
The calibrators of the TP Premium series are characterised by their unparalleled performance and outstanding operating comfort. Employing the intuitive menu structure, all necessary inputs can be made quickly and easily. The large touch screen has plenty of room to display the reference, target, and devices under test temperatures.
At the end of a calibration process, the TP Premium provides the complete calibration certificate. The continuously growing bandwidth of supported temperature ranges supports an increasing number of temperature sensors on the market. They can be calibrated with a resolution of up to 0.001 ℃ / K and thus meet the highest requirements, e.g. of the food and pharmaceuticals industry.
- ✓ Patented control technology - time-saving up to 50 %
- ✓ Fastest stabilisation times on the market
- ✓ Patented touchscreen function
Whether for on-site use in the workshop, in the measuring and test room, or laboratory, SIKA temperature calibrators can be used everywhere. A wide range of industries and various application areas are covered.
- • Assembly, commissioning
- • Manufacturing, production
- • Maintenance, service
- • Quality assurance, test equipment monitoring
- • Repair
Dry Block Function
The dry block function was developed to easily calibrate the temperature in lab and field applications. The optimum thermal coupling from block to DUT is achieved by the correct calibration insert. Ideally, the inner diameter of the borehole in the calibration insert is 0.5 mm larger than the outer diameter of the DUT. With the help of the calibration insert almost all straight temperature sensors with different lengths and diameters can be calibrated. The dry block covers the entire temperature range without changing the calibration medium. The user does not have to consider viscosities, flash points, or outgassing.
Dark-radiation Function
A patented infrared calibration insert is used to calibrate IR pyrometers or thermal imaging cameras. The special surface structure and the asymmetrical shapes form a "black cavity radiator" with an emission factor of 0.9994, and preventing the reflection of interfering radiation and emitting the desired temperature in an ideal form. The pyrometer is simply held at the prescribed distance above the measuring port and thus forms the desired measuring spot for the calibration to be carried out on the ground. The direct mounting of a tripod is possible.
Surface Temperature Function
Surface temperature sensors are calibrated employing special calibration inserts which are placed vertically with the required contact pressure. The best possible temperature reference point on the surface of the calibration insert is found by switching the calibrator control to the external reference temperature sensor. This is located directly under the front face of the calibration insert. The calibration insert is designed in such a way that the best temperature homogeneity is achieved in the middle of the front face. Good thermal contact is enabled by the special surface structure of the front face. The use of heat-conducting paste or other heat transfer media is not necessary.
Accessories
The extensive range of accessories for the TP Premium series enables time-saving calibration setups. For example, DUT can be identified via a barcode scanner or an ongoing or already completed calibration process can be displayed via a WLAN connection on a PC or smartphone anywhere in the world. Thanks to the possibility of connecting an external measuring bridge directly to the calibrators, many DUTs can be calibrated at the same time - including the automatic generation of individual SIKA works calibration certificates.
Please note that the price quoted is for a basic model without any customised options, please contact us to customise your apparatus.
TP 37200E.2 / TP 37165E.2 Linearisation Example: EPLIKDBE000 | |||
---|---|---|---|
With Linearisation (min. 1) | EPLIKDB | ||
Reference sensor | |||
Internal | I00 | ||
External | E00 | ||
Linearisation Points | |||
Standard | 0 | ||
Custom Specific | K |
TP 37450E.2 Linearisation Example: EPLIKDBE000 | ||||
---|---|---|---|---|
With Linearisation (min. 1) | EPLIK | |||
Function | ||||
Dry block (Air Shield Insert)* | AS | |||
Dry block | DB | |||
Infrared | IR | |||
Surface calibration* | SU | |||
Reference sensor | ||||
Internal | I00 | |||
External | E00 | |||
Linearisation Points | ||||
Standard | 0 | |||
Custom Specific | K |
*External reference is needed.
TP 37700E.2 Linearisation Example: EPLIKDBE000 | |||
---|---|---|---|
With Linearisation (min. 1) | EPLIK | ||
Function | |||
Dry block external (Air Shield Insert) | ASE00 | ||
Dry block internal | DBI00 | ||
Linearisation Points | |||
Standard | 0 | ||
Custom Specific | K |
Datasheets
Manuals & Guides
Certifications & Declarations
TP 37200E.2 Technical Data
Dry Block
Temperature range | -55...200 ℃ at ambient temperature 20 ℃ | |
---|---|---|
Dimension of the calibration insert | ø28 × 150 mm (calibration insert easily exchangeable) | |
Dry block | External reference temperature sensor | Internal reference temperature sensor |
Display accuracy | ±0.27 ℃ | ±0.34 ℃ |
Temperature stability | ±0.003 ℃ | ±0.020 ℃ |
Temperature distribution | ||
→ Axial | ±0.250 ℃ | — |
→ Radial | ±0.070 ℃ | — |
Influence of load | ±0.070 ℃ | ±0.220 ℃ |
Stabilisation time (with external reference temperature sensor) |
||
→ to ±0.05℃ | From 1 min | |
→ to ±0.005℃ | From 5 min | |
Heating time | ||
→ 20 ℃...200 ℃ | 9 min | |
→ -55 ℃...200 ℃ | 12 min | |
Cooling time | ||
→ 20 ℃...-55 ℃ | 35 min | |
→ 200 ℃...20 ℃ | 18 min | |
Resolution of the temperature display | 0.1 / 0.01 / 0.001 ℃ (selectable) | |
Hysteresis | ±0.010 ℃ | |
Temperature units | ℃ / ℉ / K (selectable) | |
Reference temperature sensor | Internal, fixed installation / external (selectable) | |
Interfaces | Ethernet, 3 x USB | |
Connectivity | OPC UA, serial communication, HTTP. Details and further possibilities on request. | |
Dimensions | ||
→ Width | 210 mm | |
→ Height | 380 + 50 mm (Handle) | |
→ Depth | 300 mm | |
Weight | Approx. 15 kg | |
Power supply | 100...240 VAC, 50 / 60 Hz | |
Power consumption | Approx. 555 W | |
Adjustable temperature range | -60...200 ℃ | |
Display | Brilliant color touchscreen (7 inches), multi-panel safety glass |
TP 37165E.2 Technical Data
Dry Block
Temperature range | -35...165 ℃ at ambient temperature 20 ℃ | |
---|---|---|
Dimension of the calibration insert | ø28 × 150 mm (calibration insert easily exchangeable) | |
Dry block | External reference temperature sensor | Internal reference temperature sensor |
Display accuracy | ±0.21 ℃ | ±0.23 ℃ |
Temperature stability | ±0.003 ℃ | ±0.010 ℃ |
Temperature distribution | ||
→ Axial | ±0.200 ℃ | — |
→ Radial | ±0.040 ℃ | — |
Influence of load | ±0.040 ℃ | ±0.100 ℃ |
Stabilisation time (with external reference temperature sensor) |
||
→ to ±0.05℃ | From 1 min | |
→ to ±0.005℃ | From 5 min | |
Heating time | ||
→ 20 ℃...165 ℃ | 14 min | |
→ -35 ℃...165 ℃ | 16 min | |
Cooling time | ||
→ 20 ℃...-30 ℃ | 13 min | |
→ 165 ℃...20 ℃ | 11 min | |
Resolution of the temperature display | 0.1 / 0.01 / 0.001 ℃ (selectable) | |
Hysteresis | ±0.010 ℃ | |
Temperature units | ℃ / ℉ / K (selectable) | |
Reference temperature sensor | Internal, fixed installation / external (selectable) | |
Interfaces | Ethernet, 3 x USB | |
Connectivity | OPC UA, serial communication, HTTP. Details and further possibilities on request. | |
Dimensions | ||
→ Width | 210 mm | |
→ Height | 380 + 50 mm (Handle) | |
→ Depth | 300 mm | |
Weight | Approx. 13.5 kg | |
Power supply | 100...240 VAC, 50 / 60 Hz | |
Power consumption | Approx. 375 W | |
Adjustable temperature range | -50...165 ℃ | |
Display | Brilliant color touchscreen (7 inches), multi-panel safety glass |
TP 37450E.2 Technical Data
Dry Block, Dry Block Air Shield Insert, Infrared Calibration & Surface Calibration
Temperature range | Room temperature...450 ℃ | |
---|---|---|
Dimension of the calibration insert | ø60 × 150 mm (calibration insert easily exchangeable) | |
Dry block Air Shield Insert | External reference temperature sensor | |
Display accuracy | ±0.2 ℃ | |
Temperature stability | ±0.010 ℃ | |
Temperature distribution | ||
→ Axial | ±0.100 ℃ | |
→ Radial | ±0.015 ℃ | |
Influence of load | ±0.010 ℃ | |
Dry block | External reference temperature sensor | Internal reference temperature sensor |
Display accuracy | ±0.25 ℃ | ±0.3 ℃ |
Temperature stability | ±0.010 ℃ | ±0.030 ℃ |
Temperature distribution | ||
→ Axial | ±0.300 ℃ | ±0.300 ℃ |
→ Radial | ±0.060 ℃ | ±0.060 ℃ |
Influence of load | ±0.015 ℃ | ±0.015 ℃ |
Infrared calibration | External reference temperature sensor | Internal reference temperature sensor |
Display accuracy | ±0.5 ℃ | ±0.5 ℃ |
Temperature stability | ±0.030 ℃ | ±0.100 ℃ |
Emission factor | 0.9994 | |
Surface calibration | External reference temperature sensor | |
Display accuracy | ±1 ℃ | |
Temperature stability | ±0.250 ℃ | |
Stabilisation time (with external reference temperature sensor) |
||
→ to ±0.05℃ | From 1 min | |
→ to ±0.005℃ | From 5 min | |
Heating time | ||
→ 20 ℃...440 ℃ | 17 min | |
Cooling time | ||
→ 450 ℃...30 ℃ | 36 min | |
Resolution of the temperature display | 0.1 / 0.01 / 0.001 ℃ (selectable) | |
Hysteresis | ±0.010 ℃ | |
Temperature units | ℃ / ℉ / K (selectable) | |
Reference temperature sensor | Internal / external (selectable) | |
Interfaces | Ethernet, 3 x USB | |
Connectivity | OPC UA, serial communication, HTTP. Details and further possibilities on request. | |
Dimensions | ||
→ Width | 210 mm | |
→ Height | 330 + 50 mm (Handle) | |
→ Depth | 300 mm | |
Weight | Approx. 11 kg | |
Power supply | 110...115 V 60 Hz / 230 V 50 Hz Protective conductor (PE) needed | |
Power consumption | Approx. 1000 W | |
Adjustable temperature range | 0...450 ℃ | |
Display | Brilliant color touchscreen (7 inches), multi-panel safety glass |
TP 37700E.2 Technical Data
Dry Block & Dry Block Air Shield Insert
The TP 37700 can be operated up to 700 ℃ (1292 ℉). For physical reasons, it achieves the best accuracy at temperatures up to 660 ℃ (1220 ℉). For temperatures between 660 (1220 ℉) and 700 ℃ (1292 ℉), we recommend the use of a separate reference thermometer.
Temperature range | Room temperature...700 ℃ |
---|---|
Dimension of the calibration insert | ø29 × 150 mm (calibration insert easily exchangeable) |
Dry block Air Shield Insert All values determined at 660℃ |
External reference temperature sensor |
Display accuracy | ±0.27 ℃ |
Temperature stability | ±0.015 ℃ |
Temperature distribution | |
→ Axial | ±0.400 ℃ |
→ Radial | ±0.020 ℃ |
Influence of load | ±0.020 ℃ |
Dry block All values determined at 660℃ |
Internal reference temperature sensor |
Display accuracy | ±0.43 ℃ |
Temperature stability | ±0.100 ℃ |
Temperature distribution | |
→ Axial | ±0.400 ℃ |
→ Radial | ±0.040 ℃ |
Influence of load | ±0.180 ℃ |
Stabilisation time (with external reference temperature sensor) |
|
→ to ±0.05℃ | From 1 min |
→ to ±0.005℃ | From 5 min |
Heating time | |
→ 20 ℃...690 ℃ | 19 min |
Cooling time | |
→ 700 ℃...30 ℃ | 85 min |
Resolution of the temperature display | 0.1 / 0.01 / 0.001 ℃ (selectable) |
Hysteresis | ±0.015 ℃ |
Temperature units | ℃ / ℉ / K (selectable) |
Reference temperature sensor | Internal / external (selectable) |
Interfaces | Ethernet, 3 x USB |
Connectivity | OPC UA, serial communication, HTTP. Details and further possibilities on request. |
Dimensions | |
→ Width | 210 mm |
→ Height | 330 + 50 mm (Handle) |
→ Depth | 300 mm |
Weight | Approx. 10 kg |
Power supply | 110...115 V 60 Hz / 230 V 50 Hz Protective conductor (PE) needed |
Power consumption | Approx. 1000 W |
Adjustable temperature range | 0...700 ℃ |
Display | Brilliant color touchscreen (7 inches), multi-panel safety glass |
Reviews
SKU | TP 37 |
---|