Seaward Clare HAL101 AC/DC Hipot & DC Insulation Tester

The entry-level model in Seaward Clare's HAL series, the HAL101 is a simple dielectric strength and DCIR tester suitable for testing the electrical insulation performance of electrical equipment.

The HAL101 is best suited for applications where you are measuring items one at a time. Once connected to the item you need to test, the HAL101 can be used to perform AC and DC flash/hipot tests, DC insulation resistance tests and also has an ARC detection mode.

All results can be stored internally on the HAL101's memory, allowing you to easily recall testing data at a later date. Built-in ports can be also used to connect optional barcode scanners and printers to the instrument, speeding up asset management and also allowing quick, seamless printing of your testing results (please note that both the barcode scanner and printer are sold separately).

HAL Series Production Line Safety Testers

In the HAL101's internal memory, up to 6000 testing results, including details of products, user, location, site, date and time can all be stored, ready for transfer to PC for further analysis.

The HAL101 also has fully isolated test outputs which are in compliance with EN50191.

Seamlessly Works with Seaward Safety e-Base Production Line Safety Software*

*Please note that this software package is sold separately

Take complete control over your HAL101 with Safety e-Base software for your PC.

With this software, you'll be able to create totally customisable test sequences which can be fully operated via a PC from a safe, convenient location, even if that's away from the production line.

A customisable user interface allows you to provide user prompts with texts and pictures (ideal for training new operators) and you can be guided step-by-step through the testing process, reducing human error and ensuring higher levels of productivity.

e-Base also has the added benefit of totally traceable results with virtually unlimited storage on your PC or network drive.

Features include:

  • Reduce human error and increase efficiency with PC remote control
  • Programme test sequences to start on customisable trigger commands
  • Customise test sequences for different products and standards - ensure your test system will withstand the test of time
  • Improve traceability and auditability - store test results automatically to your PC
  • Programme user prompts in text/picture format to make the process easier for new operators

Seaward Clare HAL101 AC/DC Hipot and DC Insulation Tester Key Features

  • Highly precise, easy-to-use instrument for testing the electrical safety of a single item
  • Insulation resistance test with variable DC output voltage from 0.10kV to 6.00kV
  • AC/DC dielectric strength testing with programmable voltage range from 0.10kV to 5.00kV (AC)/ 6.00kV (DC)
  • Store up to 6000 test results in non-voltatile flash memory - includes space for information on product, user, location, site, date and time
  • Highly accurate leakage and DCIR measurements
  • Large-size, full graphical display with real-time clock
  • Connects to a PC, barcode scanner and printer (requires software for use on PC - printers/barcode scanners sold separately)
  • Seamlessly works with Seaward's optional Safety e-Base software for full PC control
  • Automatic test code generation
  • Fast startup
  • Fully isolated test ouputs in compliance with EN50191

What's Included?

  • Seaward Clare HAL101 Tester
  • Instruction Manual
  • Power Cord
  • Guard Plug
  • Comms Cable
  • Test Probes/Clips

Optional Accessories

The HAL101 has a wide range of optional accessories available - please see below.

  • 194A922 - CCD Barcode Scanner
  • Clare Safety e-Base Pro Software
  • H-5009 - Class I Output Socket Box
  • H-5014 - Data Cable
  • Desk Test n Tag Printer
  • H-5030 - Desk Test n Tag Printer Lead
  • 01521/1 - Earth Test Clip
  • 01520/1 - Earth Test Probe
  • Y160H - Fault Simulator
  • Guard Plug with Link
  • H-5003 - Hipot Test Clip (Black)
  • H-5003/R - Hipot Test Clip (Red)
  • 03918/2 - Hipot Test Probe (Red)
  • PAT Checkbox
  • PATGuard 3 PAT Software
  • 483A910 - PowerSmart Module - increases single phase load testing, current up to 50A
  • Status Beacon
  • SwitchSmart

Seaward HAL Series Comparison Chart

There are multiple models in the Seaward HAL series. This comparison chart shows the difference between each.

Features HAL101 HAL102 HAL103 HAL104 HAL LED
Earth/Ground Bond - -
AC Flash/Hipot
DC Flash/Hipot
DC Insulation Resistance
ARC Detection
Leakage - - -
Load Power - - - -
Load Power (200mVA resolution) - - - -
Power Factor - - -
Results Memory
Automation Option
Barcode Scanner/Printer Option
Internal Scanner - - - -

Seaward Clare HAL101 Technical Specifications

Insulation Resistance Test
DC Output Voltage 0.10kV to 6.00kV
(10V/step when manually adjusted)
(50V/step in test sequence program)
Display Range 0.01MΩ to 500MΩ
Display Range/Accuracy 0.03MΩ to 350MΩ: ±5% ± 5 counts
350MΩ to 500MΩ: indication only
Display Resolution 0.01MΩ
PASS/FAIL Level 0.00MΩ to 500.0MΩ
AC Dielectric Strength Test
Programmable Voltage Range 0.10 to 5.00kV
(10Vstep when manually adjusted)
50V/step in test sequence program)
Frequency (Independent of Supply) 50 or 60Hz
Voltage Display Range/Accuracy 0.10kV to 5.00kV: ±1% ±5 counts
Voltage Display Resolution 0.01kV
Current Display Range/Accuracy 0.01mA to 20.00mA: ±1% ± 5 counts
Current Display Resolution 0.01mA
Selectable Range of PASS/FAIL Levels 0.01mA to 20.00mA
Maximum Current Output 20.00mA @ 5kV
Optional Arc Detection 9 levels
DC Dielectric Strength Test
Programmable Voltage Range 0.10 to 6.00kV
(10V/step when manually adjusted)
(50V/step in test sequence program)
Voltage Display Range and Accuracy 0.10 to 6.00kV: ±1% ±5 counts
Voltage Display Resolution 0.01kV
Current Display Range and Accuracy 0.01mA to 10.00mA: ±1% ±5 counts
Current Display Resolution 0.01mA
Selectable Range of PASS/FAIL Levels 0.01mA to 10.00mA
Maximum Current Output 10.00mA
Optional Arc Detection 9 levels
Leakage Current Measurement
Display Range and Accuracy 0.10mA to 20.00mA ±1% ±5 counts
Display Resolution 0.01mA
Selectable Range of PASS/FAIL Levels 0.01mA to 20mA
Touch Current Measurement
Display Range and Accuracy 0.02mA to 5.00mA ±1% ±5 counts
Display Resolution 0.01mA
Selectable Range of PASS/FAIL Levels 0.02mA to 5.00mA
Measuring Device IEC/EN 60990 Fig 4
IEC/EN 60990 Fig 3 or Fig 5 on request
General Specifications
Case Dimensions 200 x 300 x 370mm (W x H x D)
Weight 15kf approx
Operating Voltage Range Selectable: 115 to 230V AC RMS, 50/60Hz
EN50191 Compliant when used with appropriate accessories
Accessories related to this product:
Seaward Clare HAL101 Hipot Tester - Datasheet Datasheets Seaward Clare HAL101 Hipot Tester - User Manual Manuals & Guides

Reviews

Additional Information
SKU H101
Manufacturer Part Code H101
Country of Origin United Kingdom