The entry-level model in Seaward Clare's HAL series, the HAL101 is a simple dielectric strength and DCIR tester suitable for testing the electrical insulation performance of electrical equipment.
The HAL101 is best suited for applications where you are measuring items one at a time. Once connected to the item you need to test, the HAL101 can be used to perform AC and DC flash/hipot tests, DC insulation resistance tests and also has an ARC detection mode.
All results can be stored internally on the HAL101's memory, allowing you to easily recall testing data at a later date. Built-in ports can be also used to connect optional barcode scanners and printers to the instrument, speeding up asset management and also allowing quick, seamless printing of your testing results (please note that both the barcode scanner and printer are sold separately).
In the HAL101's internal memory, up to 6000 testing results, including details of products, user, location, site, date and time can all be stored, ready for transfer to PC for further analysis.
The HAL101 also has fully isolated test outputs which are in compliance with EN50191.
Seamlessly Works with Seaward Safety e-Base Production Line Safety Software*
*Please note that this software package is sold separately
Take complete control over your HAL101 with Safety e-Base software for your PC.
With this software, you'll be able to create totally customisable test sequences which can be fully operated via a PC from a safe, convenient location, even if that's away from the production line.
A customisable user interface allows you to provide user prompts with texts and pictures (ideal for training new operators) and you can be guided step-by-step through the testing process, reducing human error and ensuring higher levels of productivity.
e-Base also has the added benefit of totally traceable results with virtually unlimited storage on your PC or network drive.
Features include:
- Reduce human error and increase efficiency with PC remote control
- Programme test sequences to start on customisable trigger commands
- Customise test sequences for different products and standards - ensure your test system will withstand the test of time
- Improve traceability and auditability - store test results automatically to your PC
- Programme user prompts in text/picture format to make the process easier for new operators
Seaward Clare HAL101 AC/DC Hipot and DC Insulation Tester Key Features
- Highly precise, easy-to-use instrument for testing the electrical safety of a single item
- Insulation resistance test with variable DC output voltage from 0.10kV to 6.00kV
- AC/DC dielectric strength testing with programmable voltage range from 0.10kV to 5.00kV (AC)/ 6.00kV (DC)
- Store up to 6000 test results in non-voltatile flash memory - includes space for information on product, user, location, site, date and time
- Highly accurate leakage and DCIR measurements
- Large-size, full graphical display with real-time clock
- Connects to a PC, barcode scanner and printer (requires software for use on PC - printers/barcode scanners sold separately)
- Seamlessly works with Seaward's optional Safety e-Base software for full PC control
- Automatic test code generation
- Fast startup
- Fully isolated test ouputs in compliance with EN50191
What's Included?
- Seaward Clare HAL101 Tester
- Instruction Manual
- Power Cord
- Guard Plug
- Comms Cable
- Test Probes/Clips
Optional Accessories
The HAL101 has a wide range of optional accessories available - please see below.
- 194A922 - CCD Barcode Scanner
- Clare Safety e-Base Pro Software
- H-5009 - Class I Output Socket Box
- H-5014 - Data Cable
- Desk Test n Tag Printer
- H-5030 - Desk Test n Tag Printer Lead
- 01521/1 - Earth Test Clip
- 01520/1 - Earth Test Probe
- Y160H - Fault Simulator
- Guard Plug with Link
- H-5003 - Hipot Test Clip (Black)
- H-5003/R - Hipot Test Clip (Red)
- 03918/2 - Hipot Test Probe (Red)
- PAT Checkbox
- PATGuard 3 PAT Software
- 483A910 - PowerSmart Module - increases single phase load testing, current up to 50A
- Status Beacon
- SwitchSmart
Seaward HAL Series Comparison Chart
There are multiple models in the Seaward HAL series. This comparison chart shows the difference between each.
Features | HAL101 | HAL102 | HAL103 | HAL104 | HAL LED |
---|---|---|---|---|---|
Earth/Ground Bond | - | - | √ | √ | √ |
AC Flash/Hipot | √ | √ | √ | √ | √ |
DC Flash/Hipot | √ | √ | √ | √ | √ |
DC Insulation Resistance | √ | √ | √ | √ | √ |
ARC Detection | √ | √ | √ | √ | √ |
Leakage | - | - | - | √ | √ |
Load Power | - | - | - | √ | - |
Load Power (200mVA resolution) | - | - | - | - | √ |
Power Factor | - | - | - | √ | √ |
Results Memory | √ | √ | √ | √ | √ |
Automation Option | √ | √ | √ | √ | √ |
Barcode Scanner/Printer Option | √ | √ | √ | √ | √ |
Internal Scanner | - | √ | - | - | - |
Seaward Clare HAL101 Technical Specifications
Insulation Resistance Test | |
---|---|
DC Output Voltage | 0.10kV to 6.00kV (10V/step when manually adjusted) (50V/step in test sequence program) |
Display Range | 0.01MΩ to 500MΩ |
Display Range/Accuracy | 0.03MΩ to 350MΩ: ±5% ± 5 counts 350MΩ to 500MΩ: indication only |
Display Resolution | 0.01MΩ |
PASS/FAIL Level | 0.00MΩ to 500.0MΩ |
AC Dielectric Strength Test | |
Programmable Voltage Range | 0.10 to 5.00kV (10Vstep when manually adjusted) 50V/step in test sequence program) |
Frequency (Independent of Supply) | 50 or 60Hz |
Voltage Display Range/Accuracy | 0.10kV to 5.00kV: ±1% ±5 counts |
Voltage Display Resolution | 0.01kV |
Current Display Range/Accuracy | 0.01mA to 20.00mA: ±1% ± 5 counts |
Current Display Resolution | 0.01mA |
Selectable Range of PASS/FAIL Levels | 0.01mA to 20.00mA |
Maximum Current Output | 20.00mA @ 5kV |
Optional Arc Detection | 9 levels |
DC Dielectric Strength Test | |
Programmable Voltage Range | 0.10 to 6.00kV (10V/step when manually adjusted) (50V/step in test sequence program) |
Voltage Display Range and Accuracy | 0.10 to 6.00kV: ±1% ±5 counts |
Voltage Display Resolution | 0.01kV |
Current Display Range and Accuracy | 0.01mA to 10.00mA: ±1% ±5 counts |
Current Display Resolution | 0.01mA |
Selectable Range of PASS/FAIL Levels | 0.01mA to 10.00mA |
Maximum Current Output | 10.00mA |
Optional Arc Detection | 9 levels |
Leakage Current Measurement | |
Display Range and Accuracy | 0.10mA to 20.00mA ±1% ±5 counts |
Display Resolution | 0.01mA |
Selectable Range of PASS/FAIL Levels | 0.01mA to 20mA |
Touch Current Measurement | |
Display Range and Accuracy | 0.02mA to 5.00mA ±1% ±5 counts |
Display Resolution | 0.01mA |
Selectable Range of PASS/FAIL Levels | 0.02mA to 5.00mA |
Measuring Device | IEC/EN 60990 Fig 4 IEC/EN 60990 Fig 3 or Fig 5 on request |
General Specifications | |
Case Dimensions | 200 x 300 x 370mm (W x H x D) |
Weight | 15kf approx |
Operating Voltage Range | Selectable: 115 to 230V AC RMS, 50/60Hz |
EN50191 | Compliant when used with appropriate accessories |