All Peak Electronics Products

Peak Electronics LCR45 LCR and Impedance Meter

SKU
LCR45
Rating:
100%
(1)
Reward Points
:
75
£90.00 £75.00
3 in Stock
Advice & Support Call 01642 931329 or contact us
Price Match We won't be beaten, let us know
Peak Electronics LCR45 LCR and Impedance Meter

An upgraded version of Peak Electronic's LCR40, the LCR45 uses finer measurement resolution and advanced mathematical analysis to give users the ability to carry out comprehensive analysis of passive components (inductors, capacitors and resistors).

To use, simply choose between the available auto/manual modes and connect the test leads to the component you wish to test. Within seconds the LCR45 will identify the type of component, display the component's value and adjust itself constantly over a time period. By scrolling through a selection of pages it's also possible see more information about other component parameters such as complex impedance, complex admittance or magnitude and phase of impedance.

Designed with simplicity in mind, the LCR45 is incredibly compact, features just two buttons for operation and is equipped with 2mm compatible detachable micro-hooks, allowing connectivity to 2mm compatible probes.

Peak Electronics LCR45 Passive Component Impedance Meter Key Features

  • Detailed analysis of components with information on component summary, complex impedance and magnitude/phase
  • Supplied with gold-plated removable hook probes
  • Fluid measurements with hold function
  • Automatic or manual component type selection
  • Automatic or manual test frequency (DC, 1kHz, 15kHz or 200kHz)
  • Enhance measurements resolution (0.2uH, 0.2pF and 0.2Ω)
  • Easy menu system for access to user settings
  • Enhanced compensation for component parasitics and losses (e.g core losses)
  • Automatic or manual power-off

Peak Electronics LCR45 Technical Specifications

Parameter  Minimum   Typical   Maximum   Notes 
Resistance Range   2MΩ  
Resolution 0.1Ω 0.2Ω    
Accuracy Typically ±1.0%±0.6Ω 1,2,6
Capacitance Range 0pF   10000μF  
Resolution 0.1pF 0.2pF    
Accuracy Typically ±1.5% ±0.6pF 1,2,5
Inductance Range 0μH   10H  
Resolution 0.1μH 0.2μH    
Accuracy Typically ±1.5%±0.6μH 1,2,4
Passive

Component

Impedance
Re and Im Typically ±1.5%±10LSD 4,5,6
Magnitude Typically ±1.5%±10LSD 4,5,6
Phase Typically±5° 4,5,6
Measurement Sample Rate 0.5Hz 1.5Hz 2Hz  
Peak Test Voltage (across O/C) -1.05V   +1.05V  
 Peak Test Current (through S/C)  -3.25mA   +3.25mA  
Test Frequency

Accuracy
1kHz Typically ±0.5%  
14.9254kHz 7
200kHz  
Sine Purity Tyipcally -60dB 3rd Harmonic
Operating Temperature Range 10°C   40°C 3
Battery Operating Voltage 8.5V   13V  


1. Within 12 months of factory calibration. Please contact us if you require a full re-calibration and/or certification of traceable calibration.
2. Specified at temperatures between 15°C and 30°C.
3. Subject to acceptable LCD visibility.
4. For inductances between 100µH and 100mH in fully auto mode.
5. For capacitances between 200pF and 500nF in fully auto mode.
6. For resistances between 10Ω and 1MΩ in fully auto mode.
7. For clarity, the 14.9254kHz test frequency is shown as 15kHz.

Peak Electronics LCR45 LCR and Impedance Meter - Datasheet Datasheets Peak Electronics LCR45 LCR and Impedance Meter - User Guide Manuals & Guides

Reviews

Additional Information
SKU LCR45
Manufacturer Part Code LCR45