Free Beanie Hat. Spend £250+ (ex VAT) View
All PASS Products

PASS PPC-24 Process Calibrator / Volt mA Calibrator

SKU
PPC-24
Rating:
0%
(0)
Reward Points
:
419
£502.80 £419.00
In Stock
Free UK Delivery*
Advice & Support Call 01642 931329 or contact us
Price Match We won't be beaten, let us know
PASS PPC-24 Process Calibrator / Volt mA Calibrator

PASS PPC-24 Process Calibrator / Volt mA Calibrator

The PASS PPC-24 Process Calibrator is a compact device designed for measuring or outputting 0-24 mA DC loops and 0-20 V DC voltage. Its wide screen ensures comfortable reading.

The PPC-24 is as a versatile instrument for measuring and providing a DC loop ranging from 0 to 24 mA with an impressive accuracy of 0.015%. Additionally, it can handle DC voltage measurements within the 0-20 V range, maintaining an accuracy of approximately +/- 0.02%.

An advantage of this type of instrument is its ability to calibrate without removing the target instrument from the process. It's also preferable to calibrate an instrument in its actual operating environment. However, not all quantities can be handled this way. Currently, on-site calibrators are primarily used for electrical signals and temperature.

The calibrator serves two primary functions. First, it simulates or generates precisely defined values for a signal, which are applied upstream in the measurement chain. Second, it measures the signal. If a panel meter is present at the chain's end, you can compare its value with the calibrator's indication. Without a panel meter, the simulated input value is compared directly to the signal measured by the calibrator.

When operating independently on battery power, the process calibrator provides an extended battery life for field use. When connected to the mains, it can be used continuously without any time restrictions.

Key Features

  • Loop Measurement
  • DC mA Measurement
  • DC Voltage Output
  • DC V Measurement
  • DC mA Output Sourcing mA
  • Simulate a Transmitter
  • Auto Shut Off
  • Display all Symbols

What's Included?

  • PASS PPC-24 Voltage/Current Process-Signal Calibrator
  • Carry Case
  • Pair of Test Leads and Clips
  • 6x AAA 1.5 V Batteries
  • Power Adaptor
  • Operating Manual
Input/Output Voltage Parameters
Function Range Resolution
DC V mV Input 0 ~ 100 mV 0.01 mV
0 ~ 20 V 0.001 V
DC V mV Output 0 ~ 100 mV 0.01 mV
0 ~ 20 V 0.001 V
Loop Power Output 24V DC N/A
Input/Output mA Parameters
DC mA Input 0 ~ 24 mA 0.001 mA
DC mA Output 0 ~ 24 mA 0.001 mA
DC V Input/Output
Function Range Accuracy±
(% of reading + Counts)
100 mV 0.01 mV 0.02 % + 3
20 V 0.001 V 0.02 % + 3
Input Impedance 2MΩ (nominal), < 100pF
Over Voltage Protection 30 V
Voltage Driver Capability 1 mA
DC mA Input and Output
Function Range Accuracy±
(% of reading + Counts)
24 mA 0.001mA 0.015 % + 3
Overload Protection 125 mA, 250V fast-acting fuse
Percent Display 0%=4mA, 100%=20mA
Source Mode When output is more than 15mA @500Ω, change the power supply to external power. The maximum load is 24mA @700Ω when using external power.
Simulate Mode External loop voltage requirement: 24V nominal, 30V maximum, 12V minimum.
Loop Power 24 V ± 10%
General Specifications
Storage Temperature -40°C ~ 60°C
Operating Temperature -10°C ~ 55°C
Operating Altitude 3000 meters maximum
Temperature Coefficient ±0.005% of range per °C for the temperature range -10°C to 18°C and 28°C to 55°C
Relative Humidity 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, 35% up to 55°C
Shock Random, 2g, 5Hz to 500Hz
Safety 1 metre drop test
Power Requirements 6x AAA 1.5 V Batteries
Size (LxWxH) 204mm x 99mm x 46mm, 460g (with batteries)
30 V Max. Voltage applied between a terminal and the earth or between two terminals
PASS PPC-24 Voltage/Current Process-Signal Calibrator User Manual Manuals & Guides

Reviews

Additional Information
SKU PPC-24
Manufacturer Part Code VA700
Country of Origin China