The upgraded version of the Megger PAT320, the PAT350 includes the ability to perform flash tests. Designed for use on the desktop, the PAT350 features dedicated test buttons for easy, direct access to testing procedures. It can used in a wide variety of environments, including offices and - thanks to flash testing ability - can be used in specific areas such as manufacturing, production or tool hire shops.
The PAT350 supports all needed PAT testing procedures, including class I and class II, IEC power leads, extension leads and full tests for portable RCDs. A dedicated automatic testing mode is available for Class I and Class II testing where the PAT350 will automatically cycle through bond, insulation and operation tests. A built-in manual mode is also included for configurable tests.
Get more for your money with our exclusive PAT testing bundle packs! In these kits, you'll get everything you get as standard, plus our own essential accessories bundles which include everything you need for PAT testing.
Item Description | Megger PAT350 PAT Tester Essentials Kit (Bundle 1) | Megger PAT350 PAT Tester Professional Kit (Bundle 2) |
---|---|---|
Keep all of your equipment safe and in one place with this large-size carrying case! (appearance of case may vary depending on current stock) |
✔ | ✔ |
Need to check if your sockets are working right? Just plug this in to a standard UK plug socket and if there's a problem, the socket tester will let you know about it! |
✔ | ✔ |
Gain a certificate of competency to carry out PAT testing in just two hours with our online PAT course and exam. Teams may be trained by purchasing the course multiple times; each staff member will be assigned a unique activation code and login. |
✔ |
✔ |
Everything you need to know about PAT testing is outlined in this comprehensive book. Inside, you'll find useful information, colour-coded sections, photographic examples and much more to help you PAT test on a daily basis. |
✔ | ✔ |
Keep a record of all of your testing with our logbook. The book includes carbon-copied records, allowing you to keep one for yourself and give the other to customers or keep for a reference. |
✔ | ✔ |
When PAT testing, it's common practice to check microwaves are also not emitting high levels of radiation. With our microwave leakage detector you'll be able to do this testing with ease. |
✔ | ✔ |
You'll need this adaptor to connect your PAT tester to laptop/PC power leads and other similar equipment which has a cloverleaf-shaped connector. |
✔ | ✔ |
This adaptor is used to connect to laptop power supplies, portable projectors and other similar equipment. |
✔ | ✔ |
You'll find several appliances such as game consoles, laptop power supplies, audio/visual equipment and more use this type of connector - with this adaptor, you can connect them up to your PAT tester. |
✔ | ✔ |
A handy set of insulated screwdrivers (suitable for live working to 1000V) suitable for a wide variety of applications. |
✔ | ✔ |
You'll need these every time an appliance passes a test - just tear one off, fill in the fields (appliance ID, your name, test date and next test due date) and stick on the appliance when done. |
✔ | ✔ |
These circular-shaped PAT labels are great for plug tops and any other appliance. The labels also include space for ticking the month/year the next test is due. |
✔ | ✔ |
These passed labels wrap around a wire/cable from an appliance and are recommended for use with power supplies or on appliances where you can't stick a label. They also include space for writing your appliance ID, name, test date and next test date. |
✔ | ✔ |
You won't need these as often as your passed labels, but you'll need to apply one of these whenever an appliance fails a test. If it does fail, the item should be taken out of service for repair or replacement |
✔ | ✔ |
Keep a database of your PAT testing information on your PC and print reports with SimplyPats Manual software. Please note that this software is manual entry; you cannot download directly from the PAT tester and you will need to type in records by hand. |
✖ | ✔ |
Enjoy peace of mind with a Free PAT Card included with every purchase. Keep track of essential safety checks effortlessly!
Electrical Supply Range | |
230V AC | 230V AC ±10% |
110V AC | 110V AC ±10% |
Supply Measurement | |
Voltage | ±2% ± 1V |
Frequency | ±1% ± 0.1Hz |
Bond Test | |
Open Circuit Voltage | 9V AC ±10% ±0.1V (Supply: 230V 50Hz) |
10A Bond Test Current | 10A rms ±20% ±0.5A into 0.1Ω |
25A Bond Test Current | 25A rms ± 5% ± 0.5A into 0.1Ω |
Earth Bond Resistance Accuracy | ±5% ± 3 digits (0 to 0.49Ω) ±5% ± 5 digits (0.5 to .99Ω) |
Earth Bond Resistance Resolution | 10mΩ (0 to 1.99Ω) |
Display Range | 0 to 1.99Ω |
Bond Test Nulling | Up to 1.00Ω |
Adjustable Test Duration | User selectable from 1 sec to 20 sec |
Continuity Test | |
Note | The continuity test is a DC test performed automatically in both positive and negative directions. The average of the two results is shown |
Continuity Test Voltage | 4.0V DC -0% +10% (open circuit) |
Continuity Test Current | 200mA -0% + 10% ± 5mA (into 2Ω load) |
Continuity Resistance Accuracy | ±5% ±3 digits (0 to 0.49Ω) ±5% ±5 digits (0.5 to .99Ω) |
Continuity Resistance Resolution | 10mΩ (1 to 19.99Ω) |
Display Range | 0 to 19.99Ω |
Continuity Test Nulling | Up to 9.99Ω |
Test Duration | User selectable from 1 sec to 20 sec |
Insulation Test | |
Insulation Test Voltage | 250V DC -0%/+25% open circuit 500V DC -0%/+25% open circuit (500V DC across 0.5MΩ) |
Insulation Resistance Accuracy (230V) | ±2% ±5 digits (0 to 19.99MΩ) ±5% ± 10 digits (20 to 99.99MΩ) |
Insulation Resistance Accuracy (110V) | ±5 ±5 digits (0 to 10MΩ) |
Insulation Resistance Resolution | 0.01MΩ (0.10 to 99.99MΩ) |
Display Range | 0 to 99.99MΩ |
Test Duration | User selectable from 1 sec to 1 minutes |
Substitute Leakage Test | |
Test Voltage | 40V AC ±10% |
Test Frequency | Nominal mains 50/60Hz |
Leakage Current Accuracy | ±5% ± 5 digits |
Leakage Current Resolution | 0.01mA |
Display Range | 0 to 19.99mA |
Test Duration | User selectable from 1 sec to 1 minute Reading corrected to 230V + 10% or 110V +10% AC |
Differential Leakage Current | |
Test Voltage | Nominal mains 110V/230V AC |
Test Frequency | Nominal mains 50/60Hz |
Differential Leakage Current Accuracy | ±5% ± 5 digits |
Differential Leakage Current Resolution | 0.01mA |
Display Range | 0 to 19.99mA Reading corrected to 230V +10 % or 110V + 10% AC |
Touch Current Test | |
Test Voltage | Nominal mains 1150/230V AC |
Test Frequency | Nominal mains 50/60Hz |
Touch Current Accuracy | ±5% ±5 digits |
Touch Current Resolution | 0.01mA |
Display Range | 0 to 10mA |
Test Duration | User selectable from 1 sec to 5 sec Reading corrected to 230V + 10% or 110V + 10% AC |
Operational Test | |
Test Voltage | Nominal mains 110V/230V AC |
Accuracy | ±5% ±10 digits (0 to 99VA) ±5% ±50 digits (100 to 999VA) ±5% ±100 digits (1000 to 3700VA) |
Resolution | 1VA (0 to 3700VA) |
Display Range | 0 to 3.99KVA Reading corrected to 230V or 110V AC Results show load VA |
Extension Lead Test | |
Polartity Test Voltage | 12V |
Polarity | Leak OK Live neutral S/C Live neutral reversed Live/neutral O/C |
Flash Test | |
Flash Test Voltage | 1500V AC nominal for class 1 3000V AC nominal for class 2 |
Flash Test Current | <3.5mA short circuit @ 253V primary supply voltage |
Flash Test Breakdown Current Accuracy | ±5% ± 5 digits |
Flash Test Breakdown Current Resolution | 0.01mA |
Display Range | 0 to 3.0mA |
Test Duration | User-controlled |
Portable RCD Test | |
RCD Test Voltage | Nominal mains 110/230V |
RCD Test Frequency | 50Hz |
Test Current Accuracy | -8% to -2% (1/2 x I) ±2% to +8^ (1 x 1,5 x I) |
Trip Time Accuracy | ±1% ±5 digits |
Trip Time Resolution | 0.1ms |
Display Range | 0 to 1999ms (1/2 x I) 0 to 300ms (1 x I) 0 to 40ms (5 x I) |
Fuse Test | |
Test Voltage | 3.3V |
Warning | Audible Beep if fuse is ok |