The Megger OTS60AF is used to accurately determine the level of breakdown voltage within oil.
Featuring the ability to apply voltage up to 60kV, the OTS60AF ensures that an oil sample is safely contained within a testing chamber and that voltage can be applied without risk to the user. The breakdown of the oil can be viewed either through the transparent lid or full breakdown information is relayed through an LCD screen integrated directly into the tester.
This oil tester is suitable for use with mineral, ester and silicon insulating liquids and comes as standard with a lock-in precision oil vessel with lockable gap setting. Once an oil sample is placed within the testing chamber the OTS60AF will automatically apply voltage to the oil sample until breakdown occurs, allowing the user to easily tell what levels of voltage the insulating oil can withstand. Contaminants within oil can affect its ability to be a safe insulative property, so regular testing with a device such as the OTS60AF is essential to ensure correct oil operation.
Additional features of this tester include the ability to automatically measure the temperature of an oil sample, a full colour display that can even be viewed outdoors in direct sunlight and the ability to shut off a test and safely disperse levels of high voltage.
Megger also manufacturers identical oil testers to this one that can use test voltages up to 100kV and 80kV. We highly recommend viewing these devices by clicking a link to the left if you need a device that can apply higher voltages.
Call 01642 931 329 or complete the form:
Test Voltage | OTS60AF 0 to 60 kV rms maximum (30 kV - 0 - 30 kV) |
Voltage Resolution + Accuracy | 0.1 kV ±1% ±2 digits |
Programmed Test Sequences | ASTM D 1816-04 BS EN 60156-96 SABS EN60156 ASTM D 877A-02 CEI EN 60156-95 VDE0370 part 5 ASTM D 877B-02 IRAM 2341 AS1767.2.1 IEC 60156-95 UNE EN 6015 PA SEV EN60156 NF EN 60156 JIS C 2101-99 (M) JIS C 2101-99 (S) plus 3 custom test sequences |
Vessels 400ml (standard), 150ml (optional) | Nylon 12 chamber provides precision electrode alignment, adjustment wheels lock electrodes in position, option of 150 ml vessel for low volume oil samples |
Oil Temperature Range | 10 ºC to 65 ºC |
Oil Temperature Sensor Resolution | 1 ºC |
Power Supply | Line voltage 85 to 265 VAC Line frequency 50/60 Hz |
Interfaces | USB 2.0 compatible 2 x USB type-A (Flash drive, printer , other) 1 x USB type-B (Factory use only) |
Internal Printer (Optional) | Matrix impact printer Paper 57.5 mm wide |
External Printer | Any printer with USB interface and PCL3 driver |
Protection | Dual safety micro switches on chamber cover |
Display | 320 x 240 QVGA colour display with backlight |
Environment | |
Operating Temperature | 0 ºC to +50 ºC |
Storage Temperature | -30 ºC to +65 ºC |
Humidity | 80% RH at 40 ºC operation 95% RH at 40 ºC storage |
Altitude | 2000 m |
Safety | Designed in accordance with IEC61010 |
EMC | Light industrial IEC 61326-1 Class B, CISPR 22, CISPR 16-1 and CISPR 16-2 |
Dimensions | All models 580 mm x 420 mm x 290 mm |
Weight | All models 30 kg with printer option fitted |
Languages | English, French, German, Spanish, Chinese, Czech, Dutch, Finnish, Italian, Norwegian, Polish, Portuguese, Russian and Swedish |