The KIMO AMI310 Multifunctional Environmental Meter is an adaptable, multi-purpose instrument capable of measuring pressure, temperature, humidity, air quality, air velocity, airflow, and tachometry (depending on the probes/ modules used with it). There are seven kits available (as well as a standalone version); please see the ‘What’s Included?’ section for more details.
Up to six parameters can be measured simultaneously and saved to the KIMO AMI310 Multifunctional Environmental Meter’s internal memory. The KIMO AMI310 Multifunctional Environmental Meter can store up to 1000 datasets comprised of 20,000 data points. Its memory can be expanded using the supplied 4GB micro-SD card.
Measurements can be viewed on the KIMO AMI310 Multifunctional Environmental Meter’s large, colour display. Its rechargeable, Li-ion battery facilitates continuous measurement.
The KIMO AMI310 Multifunctional Environmental Meter can be used interchangeably with a variety of modules and probes; it includes two inputs for Pt100 temperature probes (-200°C to +600°C). Its SMART-2014 system automatically recognises wired and wireless probes. It is a versatile, customisable, and easy-to-use environmental meter.
KIMO AMI310 Multifunctional Environmental Meter
KIMO AMI310CLA Multifunctional Environmental Meter – CLA Kit
KIMO AMI310STD Multifunctional Environmental Meter – STD Kit
KIMO AMI310PRO Multifunctional Environmental Meter – PRO Kit
KIMO AMI310CRF Multifunctional Environmental Meter – CRF Kit
KIMO AMI310SRF Multifunctional Environmental Meter – SRF Kit
KIMO AMI310PRF Multifunctional Environmental Meter – PRF Kit
KIMO AMI310SK Multifunctional Environmental Meter – SK Kit
AMI310 | AMI310CLA | AMI310STD | AMI310PRO | AMI310CRF | AMI310SRF | AMI310PRF | AMI310SK | |
KIMO AMI310 Multifunctional Environmental Meter |
✔ |
✔ |
✔ |
✔ |
✔ |
✔ |
✔ |
✔ |
Ø70mm Vane Probe | ✖ |
✔ |
✖ | ✖ | ✖ | ✖ | ✖ | ✖ |
ABS Hygrometry Probe | ✖ |
✔ |
✔ |
✖ |
✖ | ✖ | ✖ | ✖ |
Hotwire Probe | ✖ |
✔ |
✔ |
✖ |
✔ |
✔ | ✖ | ✖ |
±1000 Pa Pressure Module | ✖ | ✖ |
✔ |
✖ | ✖ | ✔ | ✖ | ✖ |
Ø6mm Pitot Tube | ✖ | ✖ |
✔ |
✖ | ✖ | ✔ | ✖ | ✔ |
2x 1m Silicone Tube | ✖ | ✖ |
✔ |
✔ |
✖ | ✔ | ✔ | ✔ |
Stainless Steel Tip | ✖ | ✖ |
✔ |
✔ |
✖ | ✔ | ✔ | ✔ |
Ø100mm Vane Probe | ✖ | ✖ |
✔ |
✖ | ✖ | ✖ | ✖ | ✖ |
±500 Pa Pressure Module | ✖ | ✖ |
✖ |
✔ |
✖ | ✖ | ✔ | ✔ |
Ø100mm Telescopic Vane Probe | ✖ | ✖ |
✖ |
✔ |
✖ | ✖ | ✖ | ✖ |
Ø6mm T Pitot Tube | ✖ | ✖ |
✖ |
✔ |
✖ | ✖ | ✔ | ✖ |
Stainless Steel Hygrometry Probe | ✖ | ✖ |
✖ |
✔ |
✖ | ✖ | ✖ | ✖ |
Telescopic Hotwire Probe | ✖ | ✖ |
✖ |
✔ |
✖ | ✖ | ✔ | ✖ |
Wireless ABS Hygrometry Probe | ✖ | ✖ |
✖ |
✖ |
✔ | ✔ | ✖ | ✖ |
Wireless Ø70mm Vane Probe | ✖ | ✖ |
✖ |
✖ |
✔ | ✖ | ✖ | ✖ |
Wireless Ø100mm Vane Probe | ✖ | ✖ |
✖ |
✖ |
✖ | ✔ | ✔ | ✖ |
Wireless Stainless Steel Hygrometry Probe | ✖ | ✖ |
✖ |
✖ |
✖ | ✖ | ✔ | ✖ |
Telescopic Hotwire Probe with Gooseneck | ✖ | ✖ |
✖ |
✖ |
✖ | ✖ | ✖ | ✔ |
Connections | 2 mini-DIN connections SMART-2014 probes and 1 micro-USB port for charging and PC connection |
Power supply | Lithium-Ion battery |
Autonomy | 57 h with hygrometry probe |
Memory capacity | Up to 1000 dataset of 20 000 points in the internal memory + 4 GB micro-SD card |
Conditions of use (°C/%RH/m) |
From 0 to +50 °C. In non-condensation condition. From 0 to 2000m |
Storage temperature | From -20 to +80°C |
Auto shut-off | Adjustable from 15 to 120 minutes or Off |
Weight | 485 g |
Operating environment | Neutral gas |
European directives | 2014/30/EU ECM; 2014/35/EU Low tension; 2011/65/UE RoHS II; 2012/19/UE DEEE |
Languages | French, English, Dutch, German, Italian, Portuguese, Swedish, Norwegian, Finn, Danish, Chinese, Japanese |
Material | ABS/PC and elastomer |
Protection | IP54 |
Display | Color LCD 120 x 160 px Dimensions: 58 x 76 mm Backlight Display of 6 measurements including 3 simultaneously |
Keypad | Elastomer 10 keys |
Probe | Units | Measuring Range | Accuracy* | Resolution |
Ø14 mm vane probe | Air velocity: m/s, fpm, km/h | From 0 to 3 m/s From 3.1 to 25 m/s |
From 0.8 to 3 m/s: ±3% of reading ±0.1m/s From 3.1 to 25 m/s: ±1% of reading ±0.3 m/s |
0.1 m/s |
Airflow** : m3/h, cfm, l/s, m3/s | From 0 to 99 999 m3/h | ±3% of reading or ±0.03 *area surface (cm2 ) |
1 m3/h | |
Temperature: °C, °F | From -20 to +80 °C | ±0.4% of reading ±0.3 °C | 0.1 °C | |
Ø70 mm vane probe | Air velocity: m/s, fpm, km/h, mph | From -5 to 3 m/s From 3.1 to 35 m/s |
From 0.4 to 3 m/s: ±3% of reading ±0.1m/s From 3.1 to 35 m/s: ±1% of reading ±0.3 m/s |
0.1 m/s |
Airflow**: m3 /h, cfm, l/s, m3/s |
From 0 to 99999 m3 /h |
±3% of reading or ±0.03 *area surface (cm2) | 1 m3/h | |
Temperature: °C, °F | From -20 to +80 °C | ±0.4% of reading ±0.3°C | 0.1 °C | |
Ø100 mm vane probe | Air velocity: m/s, fpm, km/h, mph | From -5 to 3 m/s From 3.1 to 35 m/s |
From 0.3 to 3 m/s: ±3% of reading ±0.1m/s From 3.1 to 35 m/s: ±1% of reading ±0.3 m/s |
0.01 m/s 0.01 m/s |
Airflow**: m3/h, cfm, l/s, m3/s | From 0 to 99 999 m3 /h |
±3% of reading or ±0.03 *area surface (cm2) | 1 m3/h | |
Temperature: °C, °F | From -20 to +80 °C | ±0.4% of reading ±0.3°C | 0.1 °C | |
Hotwire probe | Air velocity: m/s, fpm, km/h | From 0.15 to 1 m/s From 0.15 to 3 m/s From 3.1 to 30 m/s |
±2% of reading ±0.03 m/s (Specific adjustment and calibration in option) ±3% of reading ±0.03 m/s ±3% of reading ±0.1 m/s |
0.01 m/s 0.01 m/s 0.1 m/s |
Airflow**: m3/h, cfm, l/s, m3/S | From 0 to 99 999 m3/h | ±3% of reading or ±0.03 *area surface (cm2) | 1 m3 /h |
|
Temperature: °C, °F | From -20 to +80 °C | ±0.3% of reading ±0.25°C | 0.1 °C | |
Notes | *All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation. | |||
**Calculated parameters. / ***Depending on the differential pressure element connected to the instrument. |
Pressure Module | Units | Measuring Range | Accuracy* | Resolution | Tolerated Overpressure |
MPR 500 | Pa, mmH2 O, In WG, mbar, hPa, mmHg, daPa, kPa |
From 0 to ±500 Pa From 2 to 28 m/s*** |
From -100 to +100 Pa: ±0.2% of reading ±0.8 Pa Beyond: ±0.2% of reading ±1.5 Pa |
From -100 to +100 Pa: 0.1 Pa Beyond: 1 Pa |
250 mbar |
MPR 2500 | From 0 to ±2500 Pa From 2 to 60 m/s*** |
±0.2% of reading ±2 Pa | From -100 to +100 Pa: 0.1 Pa Beyond: 1 Pa |
500 mbar | |
MPR 10000 | From 0 to ±10000 Pa From 4 to 100 m/s*** |
±0.2% of reading ±10 Pa | 1 Pa | 1200 mbar | |
MPR 500 M | mmH2 O, In WG, mbar, hPa, mmHg, daPa, kPa, PSI |
From 0 to ±500 mbar From 9 to 100 m/s*** |
±0.2% of reading ±0.5 mbar | 0.1 mbar | 2 bar |
MPR 2000 M | bar, In WG, mbar, hPa, mmHg, kPa, PSI |
From 0 to ±2000 mbar From 18 to 100 m/s*** |
±0.2% of reading ±2 mbar | 1 mbar | 6 bar |
Pitot tubes | Air velocity: m/s, fpm, km/h, mph | From 3 to 5 m/s From 5.1 to 85 m/s |
±0.3 m/s ±0.5% of reading ±0.2 m/s |
0.1 m/s | - |
Airflow**: m3 /h, cfm, l/s, m3 /s |
From 0 to 99 999 m3/h | ±0.2% of reading ±1% FS | 1 m3 /h |
- | |
Debimo blades | Air velocity: m/s, fpm, km/h, mph | From 3 to 20 m/s From 21 to 40 m/s |
±0.3 m/s ±1% of reading ±0.1 m/s |
0.1 m/s | - |
Air flow**: m3 /h, cfm, l/s, m3/s |
From 0 to 99 999 m3/h | ±0.2% of reading ±1% FS | 1 m3 /h |
- | |
Notes | *All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation | ||||
**Calculated parameters. / ***Depending on the differential pressure element connected to the instrument. |
Thermocouple | Units | Measuring range | Accuracy* | Resolution |
Thermocouple | °C, °F | K: From -200 to +1300°C J: From -100 to +750°C N: From -200 to 1300°C T: From -200 to +400°C S: From 0 to 1760°C |
K, J, N, T: S: ±0.6 °C |
0.1 °C 0.1 °C 0.1 °C 0.1 °C 0.1 °C |
Notes | *All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation.
|
|||
**Calculated parameters. / ***Depending on the differential pressure element connected to the instrument.
|
Probe | Units | Measuring Range | Accuracy* | Resolution |
Hygrometry probes SHR 110 and SHR 300 |
Relative humidity: %RH | From 3 to 98% RH | Accuracy (Repeatability, linearity, Hysteresis): ±1.5%RH (from 15°C to 25°C) Factory calibration uncertainty: ±0.88% RH Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C) |
0.1% RH |
Absolute humidity**: g/m3 | From 0 to 600 g/m3 | - | 0.1 g/m3 | |
Dew-point**: °Ctd, °Ftd | From -50 to +100 °Ctd | ±0.6% of reading ±0.5 °Ctd | 0.1°Ctd | |
Wet temperature**: °Ctw, °Ftw | From -50 to +100 °Ctw | ±0.6% of reading ±0.5 °Ctd | 0.1°Ctw | |
Enthalpy**: kj/kg | From 0 to 10 000 kj/kg | - | 0.1 kj/kg | |
Temperature: °C, °F | From -20 to +80 °C (SHR110) From -40 to +180 °C (SHR 300) |
±0.3% of reading ±0.25 °C | 0.1°C | |
Combination ratio** : g/kg | From 0 to 10 000 g/kg | - | 0.1 g/kg | |
Omnidirectional probe of airstream SOM 900 |
Air velocity: m/s, fpm, km/h | From 0.00 to 5.00 m/s | ±3% of reading ±0.05 m/s | 0.01 m/s |
Relative humidity: %RH | From 5 to 95% RH | Accuracy (Repeatability, linearity, Hysteresis): ±1.5%RH (from 15°C to 25°C) Factory calibration uncertainty: ±0.88% RH Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C) |
0.1% RH | |
Temperature: °C, °F | From -20 to +80 °C | ±0.3% of reading ±0.25 °C | 0.1 °C | |
CO2 /hygrometry/ temperature probe SCOH 112 |
Temperature: °C, °F CO2: ppm Hygrometry: %RH |
From -20 to +80 °C From 0 to 5000 ppm From 5 to 95% RH |
±0.3% of reading ±0.25 °C ±3% of reading ±50 ppm Accuracy (Repeatability, linearity, Hysteresis): ±1.5%RH (from 15°C to 25°C) Factory calibration uncertainty: ±0.88% RH Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C) |
0.1 °C 1 ppm 0.1% RH |
Notes | *All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation | |||
**Calculated parameters. / ***Depending on the differential pressure element connected to the instrument. |