The Kern ALJ-5DA Dual-Range Analytical Balance can be used to determine the weights of a variety of items. To this end, it features two ranges with upper limits of 82g and 220g, expanding its scope-for-use.
This highly accurate balance includes a number of features designed to maintain this instrument’s accuracy in a variety of environments. It features an electromagnetic force compensation function and an internal automatic adjustment feature. The latter feature is designed to maintain the balance’s accuracy in various temperatures; it will alter the balance when a temperature change of equal to or greater than 1.2°C is detected. Additionally, this feature can be configured to activate every three hours in order to verify and/or adjust the balance’s accuracy as necessary.
Other features and functions present on the Kern ALJ-5DA Dual-Range Analytical Balance include:
Recipe weights can be sent to print via the RS-232 interface and a compatible Kern printer (NOT included). In addition to this, the user is able to send selected readings to print; these weights are printed with the date and time.
Measurements are viewable via the Kern ALJ-5DA Dual-Range Analytical Balance’s LCD, backlit display with a digit height of 17mm. The user can change measurement units at the touch of a button and is able to program higher and lower alarm limits using the ergonomically designed keypad optimised for both right- and left-handed users. When either the upper or lower limits are breached the user will be alerted via an audible and visual alarm.
Adding to this instruments ergonomics is its compact size, whilst its large, glass draught shield with three sliding doors protects the items-under-investigation from contamination and provides the user with a clear view and easy access to the objects. This instrument is also supplied with a protective working cover.
The Kern ALJ-5DA Dual-Range Analytical Balance is a comprehensive weighing instrument optimised for accuracy and ease-of-use.
There are a number of Kern accessories compatible with the Kern ALJ-5DA Dual-Range Analytical Balance, including Kern calibrations. For a quick overview of the accessories available for Kern ALJ-5DA Dual-Range Analytical Balance, please see the table below.
Product Code | Product Name |
965-101 | Kern DAkkS Calibration Certificate |
KERN ALJ-A03 | Kern Draft Shield Rear Panel with Integrated Ioniser |
KERN YDB-03 | Kern Set for Density Determination |
KERN YPS-03 | Kern Weighing Table |
YBI-01A | Kern Ioniser |
YKA-08 | Kern Mains Adapter for Kern's YKN Printer |
YKN-01 | Kern Matrix Needle Printer |
YKB-A10 | Kern Thermal Receipt Rolls for Kern's YKB-01N and YKS-01 Printers |
YKB-01 | Kern Thermal Printer |
YKS-01 | Kern Statistics Printer |
YKI-01 | Kern RS-232/ Ethernet Adapter |
YKI-02 | Kern RS-232/ Bluetooth Adapter |
YKI-03 | Kern RS-232/ Wi-Fi Adapter |
YKE-01 | Kern Label Printer |
YKE-A03 | Kern Roll of Labels for Kern's YKE-01 Printer |
YKE-A02 | Kern Roll of Labels for Kern's YKE-01 Printer |
Construction | |
Casing material | Plastic |
Dimensions draft shield (W×D×H) | 190 mm x 195 mm x 225 mm |
Dimensions housing (W×D×H) | 345 mm x 215 mm x 345 mm |
Level indicator | yes |
Material weighing plate | stainless steel |
Overall dimensions mounted (W×D×H) | 210 mm x 340 mm x 330 mm |
Revolving screw feet | yes |
Weighing space (W×D×H) | 162 mm x 171 mm x 225 mm |
Weighing surface (d) | 80 mm |
Service | |
DAkkS Certificate | 963-101 |
Display | |
Display digit height | 17 mm |
Power Supply | |
Counting function | yes |
Density determination function | yes |
Peak function | yes |
Percentage determination | yes |
PRE-TARE function | yes |
Standby function | yes |
Vibration-free weighing (Animal weighing program) | yes |
Weighing with tolerance range | yes |
Measuring System | |
Adjustment options | Internal calibration (auto) |
Linearity | 0,100 mg | 0,200 mg |
Minimum weight (USP) | 80 mg |
Readability [d] | 0,010 mg | 0,100 mg |
Repeatability | 0,040 mg | 0,100 mg |
Resolution | 8.200.000 2.200.000 |
Stabilisation time under laboratory conditions: | 6 s |
Warm up time | 8 h | 8 h |
Weighing capacity [Max] | 82 g | 220 g |
Weighing system | Force compensation |
Environmental Conditions | |
Maximum humidity | 80 % |
Maximum operating temperature | 35 °C |
Minimum ambient temperature | 5 °C |
Counting | |
Counting resolution | 220000 |
Minimum piece weight at piece counting (Laboratory) | 1 mg |
Minimum piece weight at piece counting (Normal) | 10 mg |
Reference weight | yes |