or Call 01642 931 329 to discuss product options
The Isotech Isocal 6 Europa 4520 Temperature Calibrator is a multifunctional dry block calibration instrument suitable for accurately and reliably calibrating temperature sensors and probes between -45°C and 140°C. There are three versions of the Isotech Isocal 6 Europa 4520 Temperature Calibrator, Basic, Onsite and Advanced. Please see the table below for more information regarding the differences between these models.
SKU | Title | Additional Features |
IS-4520B | Isotech Isocal 6 Europa 4520B - Basic |
|
IS-4520S | Isotech Isocal 6 Europa 4520S - Onsite |
|
IS-4520 | Isotech Isocal 6 Europa 4520 - Advanced |
|
*** Please see the Comparison Chart below for more detailed information. *** |
All models support six modes of operation including Dry Block/ Metal Block Bath (default), Stirred Liquid Bath, Stirred Ice/ Water Bath, Blackbody Source, Surface Sensor Calibrator and ITS-90 Fixed Point which provides an uncertainty of down to 0.0005°C (0.5mK).
Additional accessories are required to convert this dry block calibrator into a stirred bath, blackbody, surface sensor and ITS-90 fixed point. When used as a dry block, the Isotech Isocal 6 Europa 4520 Temperature Calibrator can calibrate several thermometers simultaneously. It has a 35mm x 160mm calibration volume.
The Isotech Isocal 6 Europa 4520 Temperature Calibrator calibrates whole measurement loops and provides excellent stability, as well as distributed heating and cooling zones for good uniformity. This ensures reliable and accurate thermal performance. In addition to this, the Isotech Isocal 6 Europa 4520 Temperature Calibrator has fast heating and cooling times and is suitable for both bench and portable calibration. Please see the video below for an overview of Isotech’s 4000 Series.
Finally, the Isotech Isocal 6 Europa 4520 Temperature Calibrator meets the calibration capacity requirements of EURAMET/cg-13/v.01, “EA Guidelines on the Calibration of Temperature Block Calibrators” which was formerly EA10/13 accuracy and best performance.
The Isotech Isocal 6 Europa 4520 Temperature Calibrator is a versatile, accurate and reliable dry block temperature calibrator.
Modes of Operation
Please see the videos below for an introduction to the Isotech Isocal 6 Europa 4520 Temperature Calibrator’s six modes of operation.
Dry Block/ Metal Block Bath
Stirred Liquid Bath
Stirred Ice/ Water Bath
Blackbody Source
Surface Sensor Calibrator
ITS-90 Fixed Point
Isotech Isocal 6 Europa 4520 Temperature Calibrator Key Features
- A multifunctional dry block temperature calibrator
- Accurately and reliably calibrate temperature probes and all types of temperature sensors
- Temperature range: -45°C to +140°C
- Three models available: Basic, Onsite and Advanced (see table above)
- Six modes of operation: Dry Block/ Metal Block Bath (default), Stirred Liquid Bath, Stirred Ice/ Water Bath, Blackbody Source, Surface Sensor Calibrator and ITS-90 Fixed Point
- ITS-90 Fixed Point provides an uncertainty of down to 0.0005°C (0.5mK)
- Additional accessories are required to convert this dry-block calibrator into a stirred bath, blackbody, surface sensor and ITS-90 fixed-point calibrator
- Calibrate several thermometers simultaneously in dry block mode
- 35mm x 160mm calibration volume
- Calibrates whole measurement loops
- Provides excellent stability
- Distributed heating/ cooling zones provide good uniformity
- Accurate and reliable thermal performance
- Fast heating and cooling times
- Suitable for bench and portable calibration
- Meets calibration capacity requirements of EURAMET/cg-13/v.01, “EA Guidelines on the Calibration of Temperature Block Calibrators”
- A versatile, accurate and reliable dry block temperature calibrator
Basic, Onsite & Advanced Model Comparison Chart
Advanced | Onsite | Basic | |
Digital Display of Set and Nominal Block Temperature |
Yes | Yes | Yes |
PC Interface | Ethernet + USB Host | Serial | Serial |
Test Thermostats | Yes - Two Inputs | Yes - Single Input | No |
Independent Temperature Indicator for Reference Probe |
Yes | Yes | No |
Additional Inputs for Units Under Test | Up to 3: Two universal inputs for prt, Thermocouple or Process inputs and a further Thermocouple input | No | No |
Automatic Temperature Cycling | Yes | No | No |
Data Logging | Yes - Export to USB | No | No |
Offset Elimination | Yes - block can follow reference input |
No | No |
Choose English, French, Italian or Spanish Language |
Yes - on full colour display | No | No |
In Built Web Server | Yes | No | No |
Tamper Proof Data | Yes - Suitable for life science, automotive and aerospace applications |
No | No |
What’s Included?
- Your Choice of Isotech Isocal 6 Europa 4520 Temperature Calibrator and Corresponding Accessories
SKU | Title | What's Included? |
IS-4520B | Isotech Isocal 6 Europa 4520B - Basic |
Supply Voltage: 108/130V or 208/240v 50/60Hz - Please Specify |
IS-4520S | Isotech Isocal 6 Europa 4520S - Onsite |
Supply Voltage: 108/130V or 208/240v 50/60Hz - Please Specify |
IS-4520B | Isotech Isocal 6 Europa 4520 - Advanced |
Supply Voltage: 108/130V or 208/240v 50/60Hz - Please Specify |
Isotech Isocal 6 Europa 4520 Temperature Calibrator Technical Specifications
Temperature Range* | -45°C to 140°C |
Advanced Range | |
Stability: Dry Block / Liquid Bath | ±0.01°C |
Display Resolution | 0.001°C over whole range |
Accuracy: RTD Input Channel | ±0.05°C ±0.005% RDG |
Accuracy: Thermocouple Input Channel | E,J,K,N: ±0.2°C @ 660°C R: ±0.6°C S: ±0.7°C @ 660°C T ±0.2°C @ 150°C |
CJC Accuracy | ±0.35°C |
Basic/Site Range | |
Stability | ±0.03°C |
Display Resolution | 0.01°C from -19.99 to 99.99C then 0.1C: 0.01C Over PC Interface |
Common Specification | |
Stability | Blackbody ±0.3°C Surface Sensor ±0.5°C ITS-90 Cells ±0.0005°C |
Display Accuracy 4 | 0.15°C |
Uniformity - Between Wells Dry Block Mode (Radial) |
<0.008°C |
Uniformity - Lower 40mm (Axial) Dry Block Mode |
<0.040°C |
Uniformity - Radial Liquid Bath Mode |
<0.02°C |
Uniformity - Lower 40mm (Axial) As Liquid Bath |
<0.026°C |
Heating Time | -30°C to 140°C: 15 Mins |
Cooling Time | 140°C to 0°C: 15 Mins |
Calibration Volume | 35 x 160mm |
Standard Insert | 6 pockets, 2 x 4.5mm, 2 x 6.4mm, 1 x 8.0mm, 1 x 9.5mm diameter, all 157mm deep |
Insert Types | Choice of Three - See Accessories |
Ethernet - supporting software and USB Host | |
CJC Accuracy | 0.35°C |
Dimensions | 384H (including handle) x 212W x 312D mm |
Power | 300 Watts |
Voltage | 115Vac or 230 Vac 50/60Hz |
Weight | 14kg |
Note | (1) In ambient of 20°C: Minimum Temperature is 65°C Below Ambient, Absolute Minimum -55°C |
(2) In ambient of 20°C: Minimum Temperature is 55°C Below Ambient, Absolute Minimum -45°C | |
(3) In ambient of 20°C | |
(4) Dry Block Mode only: Comparing 4.5mm Well to Display Value |
Reviews