These Hanna meters measure low and high-range conductivity and TDS. The HI-99300 has a bottom-mounted probe, a multi-level LCD, and on-screen tutorial messages. This model is ideal for measuring low-range conductivity in µS/cm and TDS in ppm, making it a great fit for pure and tap water.
The HI-99301 is suitable for high-range conductivity in mS/cm and TDS in ppt, making it well-suited for boiler water and wastewater environments. Each meter has an IP67-rated waterproof casing and comfortably fits into the palm of your hand.
Both meters offer single-point automatic calibration with automatic temperature compensation, user-selectable EC/TDS conversion factors (0.45 to 1.00) and temperature coefficients (0.0 to 2.4%/°C). The display presents all key readings—conductivity or TDS, temperature, and calibration prompts—at a glance. The battery percentage displays at startup, and a low battery warning symbol alerts users when limited runtime remains.
Each model includes a dedicated, clog-resistant, easy-to-clean EC/TDS/temperature probe - HI-76306 for the HI-99300 and HI-763063 for the HI-99301.
HI-99300 | HI-99301 | |
---|---|---|
Range - EC | 0 to 3999 µS/cm | 0 to 20.00 mS/cm (displays mS) |
Range - TDS | 0 to 2000 ppm (mg/L) | 0.00 to 10.00 ppt |
Range - Temperature | -5.0 to 105.0°C | |
Resolution - EC | 1 µS/cm | 0.01 mS/cm |
Resolution - TDS | 1 ppm (mg/L) | 0.01 ppt (g/L) |
Resolution - Temperature | 0.1°C | |
Accuracy - EC | ±2% F.S. | |
Accuracy - TDS | ±2% F.S. | |
Accuracy - Temperature | ±0.5°C | |
TDS Factor | Adjustable, from 0.45 to 1.00 with 0.01 increments | |
Calibration | automatic, 1 point at 1382 ppm (CONV 0.5) or at 1500 ppm (CONV 0.7) or at 1413 µS/cm (others) |
automatic, 1 point at 6.44 ppt (CONV 0.5) or at 9.02 ppt (CONV 0.7) or at 12.88 mS/cm (others) |
Temperature Compensation | Automatic, with b selectable from 0.0 to 2.4 %/ °C with 0.1 increments | |
Probe | HI 763063, EC/TDS/temperature sensor, with DIN connector, 1 m (3.3') cable (included) |
|
Battery Type / Life | 3 x 1.5V AAA / approx. 500 hours of continuous use | |
Environment | 0 to 50°C (32 to 122°F); RH max 100% | |
Auto-Off | User selectable: after 8 min, 60 min or disabled | |
Casing Ingress Protection Rating | IP67 | |
Dimensions | 154 x 63 x 30 mm | |
Weight | 196g |