The Fluke P3032 Gas Deadweight Tester – Liquid Lubricated is available in five models with varying pressure ranges and measurement units. For further information regarding the differences between these models, please see the table below.
Product Code | Product Name | Pressure Range |
P3032-BAR | Fluke P3032-BAR Gas Deadweight Tester 140 bar - Liquid Lubricated | 1 to 140 bar |
P3032-KGCM2 | Fluke P3032-KGCM2 Gas Deadweight Tester 140 kgf/cm2 - Liquid Lubricated | 1 to 140 kgf/cm |
P3032-PSI | Fluke P3032-PSI Gas Deadweight Tester 2000 psi - Liquid Lubricated | 10 to 2,000 psi |
P3032-KPA | Fluke P3032-KPA Gas Deadweight Tester 14000 kPa - Liquid Lubricated | 100 to 14,000 kPa |
P3032-MPA | Fluke P3032-MPA Gas Deadweight Tester 14 MPa - Liquid Lubricated | 0.1 to 14 MPa |
The Fluke P3032 Gas Deadweight Tester can be used to create pneumatic pressure in order to calibrate pressure sensing instruments such as transmitters, transducers, gauges and pressure switches. To this end, it includes a /NPT female port to which an external gas supply may be attached, as well as a test-station design with O-rings and 1/8, 1/4, 3/8, 1/2 NPT and BSP, and M20 and M14 metric adapters, allowing users to attach instruments-under-investigation without PTFE tape or wrenches.
The high-quality needle on this tester may be used to make minor adjustments to the pneumatic pressure created. Once pressure is created the upward force causes the piston to rise in the cylinder, to counterbalance this force, accurately calibrated weight masses are added to the piston. Each weight mass is marked with a nominal pressure value relative to the high- or low-pressure piston; to determine the final pressure these values are added together. For further information regarding this process, please see the video below.
In addition to the nominal pressure values, weight masses are marked with the instrument’s serial number. The weight masses are constructed from series 3, non-magnetic, austenitic, stainless steel and are stored in a rugged weight case with a secure locking mechanism. This ensures that they may be easily transported between worksites.
The Fluke P3032 Gas Deadweight Tester’s removable lid also facilitates the easy transportation of this calibrator between worksites. Once the user has arrived at each worksite, the Fluke P3032 Gas Deadweight Tester can be easily and precisely set up due to the inclusion of adjustable feet and a mounted spirit level.
Accuracy is at the forefront of this product’s design, it is supplied with an ISO 17025 accredited calibration certificate and is calibrated to standard gravity (980.665 cm/s²). This provides the unit with a standard accuracy of 0.015% which may be improved to 0.008% via adjusting the set-up of the instrument, for instance by calibrating the instrument to the user’s local gravity. By calibrating the Fluke P3032 Gas Deadweight Tester to the user’s local gravity, the effect of this localised gravitational pull is accounted for, thereby improving the accuracy of the tester.
Also adding to the precision of the Fluke P3032 Gas Deadweight Tester, is its piston and cylinder assembly with floatation indicator which has been constructed to the highest standards and as such has certified accuracies traceable to international standards laboratories, such as NIST (National Institute of Standards and Technology). Its design ensures that the effects of distortion and thermal coefficients are lowered, whilst the instrument’s stability and durability are maintained.
Finally, the piston and cylinder assembly is oil lubricated. This ensures that when gas is introduced to the system it acts directly on the surface of the oil, providing a rate-less transfer of pressure. This design reduces the possibility of over-filling the chamber and contaminating the gas system.
PLEASE NOTE this system does not include a barrier between the oil and gas, which increases the likely hood of ‘oil-mist’ transfer. Therefore the Fluke P3032 Gas Deadweight Tester is unsuitable for oxygen service; however a version of this tester with an oxygen compatible fluid is available, please call for details.
The Fluke P3032 Gas Deadweight Tester – Liquid Lubricated is ideal for conducting accurate calibrations of pressure sensing equipment.
Fluke P3032 Gas Deadweight Tester – Liquid Lubricated Key Features
- Available in five models with varying pressure ranges and measurement units
- Ideal for calibrating pressure sensing instruments such as transmitters, transducers, gauges and pressure switches
- Create pneumatic pressure by attaching an external gas supply to the /NPT female port
- Test station design with O-rings
- 1/8, 1/4, 3/8, 1/2 NPT and BSP, and M20 and M14 metric adapters
- No need for PTFE tape or wrenches when attaching additional instruments
- High-quality needle for making fine pressure adjustments
- Piston begins to rise in the cylinder when pneumatic pressure is created
- Accurately calibrated weight masses are added to the piston to counterbalance this upward force
- Nominal pressure values relative to the high- or low-pressure piston are inscribed on each weight mass
- Final pressure is determined by adding the nominal pressure values on the weight masses together
- The instrument’s serial number is also marked on the weight masses
- Weight masses are made from series 3, non-magnetic, austenitic, stainless steel
- Weight masses may be stored in a robust weight case with a secure locking mechanism
- Removable lid
- Adjustable feet
- Mounted spirit level
- ISO 17025 accredited calibration certificate
- Calibrated to standard gravity (980.665cm/s²)
- Standard accuracy of 0.015%
- Standard accuracy may be improved to 0.008%
- Can be calibrated to local gravity
- Piston and cylinder assembly with floatation indicator
- Piston and cylinder assembly has been constructed to the highest standards
- Piston and cylinder assembly features certified accuracies traceable to international standards laboratories (NIST)
- The design of the piston and cylinder assembly limits the effects of distortion and thermal coefficients whilst maintaining the tester’s durability and stability
- Oil lubricated piston and cylinder assembly prevents the over-filling of the chamber and the contamination of the gas system
- Not suitable for oxygen service, models with an oxygen compatible fluid are available, please call for details
- Ideal for conducting accurate calibrations of pressure sensing equipment
What’s Included?
- Your Choice of Fluke P3032 Gas Deadweight Tester – Liquid Lubricated
- Weight Masses
- Removable Lid
- 1/8, 1/4, 3/8, 1/2 NPT and BSP adapters
- M20 and M14 metric adapters
- ISO 17025 Accredited Calibration Certificate
- Spare Seals
Fluke P3032 Gas Deadweight Tester – Liquid Lubricated Technical Specifications
Accuracy | |
Accuracy | ±0.015% of reading (±0.008% optional)
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Mass and PCU Materials of Construction | |
Standard weight material | Series 3 non-magnetic austenitic stainless steel |
Weight density | 7.8 g/cm³ |
Optional fractional weights | Solution heat treated aluminium |
Weight density | 2.7 g/cm³ |
Piston Material | Tungsten carbide |
Cylinder Material | Martensitic steel |
Coefficient of Expansion | 16.5 ppm/°C |
General | |
Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP |
Instrument weight | 24 lbs |
Instrument size | 17.5 W x 12 D x 8.5 H |
Mass set weight (typical) | 65 lbs |
Minimum Standard weight Increments - Pressure | 1 psi, (0.1bar) |
Operating Fluid | |
55-655 | mineral oil (standard) |
Krytox | for oxygen-safe applications (optional) |
Reviews
SKU | FLK-P3032 |
---|---|