Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert

The Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert can be used to reach temperatures 50°C below ambient (23°C), giving it a typical temperature range of -25°C to 140°C (-13°F to 284°F). It is suitable for calibrating a wide variety of temperature probes.

This instrument is available in five variations which differ only in regards to their supplied disc inserts, allowing the user to select the model most suited to his/ her needs. Each disc insert includes differently sized wells to accommodate probes with various diameters; for further information, please see the table below.

Product Code Product Name  Disc Insert Name  Disc Insert Code  Wells in Disc Insert
9103-A-256 Fluke 9103-A-256 Below Ambient Dry Well Calibrator  A 3103-2 1/16" (1.6mm), 1/4" (6.35mm), 3/8in (9.5mm), 3/16" (4.8mm), 1/8" (3.2mm), 1/2" (12.7mm)
9103-B-256 Fluke 9103-B-256 Below Ambient Dry Well Calibrator  B 3103-3 3/16" (4.8mm), 1/4" (6.35mm), 3/8" (9.5mm), 3/16" (4.8mm), 1/4" (6.35mm), 3/8" (9.5mm)
9103-C-256 Fluke 9103-C-256 Below Ambient Dry Well Calibrator  C 3103-4 6x 1/4" (6.35mm)
9103-D-256 Fluke 9103-D-256 Below Ambient Dry Well Calibrator  D 3103-6 3mm, 4mm, 6mm, 3mm, 4mm, 6mm
9103-DW-256 Fluke 9103-DW-256 Below Ambient Dry Well Calibrator  DW NO INSERT NO INSERT

The Fluke 9103-X-256 Below Ambient Dry Well Calibrator is able to match and exceed the temperature ranges of similar dry well calibrators on the market. Whereas most dry wells are capable of reaching 45°C below ambient temperature (typically equates to -20°C), the Fluke 9103-X-256 Below Ambient Dry Well Calibrator is able to reach 50°C below ambient temperature (-25°C), as mentioned above.

In addition to this, the Fluke 9103-X-256 can reach the desired temperatures quickly. For instance, from ambient (23°C) this instrument is able to reach the following temperatures in the following times:

  • 0°C – eight minutes
  • 100° -  six minutes
  • 140°C – eighteen minutes
  • -25°C – twenty minutes

This improves the instrument’s efficiency-of-use. Moreover, the Fluke 9103-X-256 Below Ambient Dry Well Calibrator is stable to ±0.02°C and has a display accuracy of ±0.25°C at all temperatures within its range.

Also adding to both the accuracy and efficiency of this instrument is its supplied NIST-traceable calibration and ability to be easily re-calibrated. In order to re-calibrate this instrument its case does not have to be opened, thereby reducing both the cost of maintenance and the instrument’s downtime; subsequently improving the user’s efficiency.

In addition to its versatility, accuracy and efficiency, the Fluke 9103-X-256 Below Ambient Dry Well Calibrator has also been optimised for practicality and ease-of-operation. To this end, it includes a front panel from which the heating and cooling rates can be adjusted and thermal switches can be checked for actuation testing.  In addition to this, the Fluke 9103-X-256 Below Ambient Dry Well Calibrator includes an RS-232 interface facilitating communication with the user’s PC.  Using this interface and the supplied Fluke Interface-It Software, the operator is able to adjust set points and ramp rates; record dry-well readings, saving the measurements to a file; collect data whilst performing a thermal switch test, and create an electronic strip chart.

Finally, the Fluke 9103-X-256 Below Ambient Dry Well Calibrator has been designed to be compact, lightweight and portable, making it suitable for applications within calibration laboratories and in-the-field.

The Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert is an efficient, comprehensive, accurate and practical temperature calibrator.

Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert Key Features

  • Suitable for calibrating a wide variety of temperature probes
  • Capable of reaching temperatures 50°C below ambient temperature (23°C)
  • Typical temperature range of -25°C to 140°C (-13°F to 284°F)
  • Available in five models with various disc inserts
  • Disc inserts allow the user to tailor the instrument to the application
  • Able to match and exceed the temperature ranges of similar dry well calibrators on the market
  • Reaches 0°C from ambient (23°C) in eight minutes
  • Reaches 100° from ambient in six minutes
  • Reaches 140°C from ambient in eighteen minutes
  • Reaches -25°C from ambient in twenty minutes
  • Stable to ±0.02°C
  • Display accuracy of ±0.25°C at all temperatures within the calibrator’s range
  • Supplied with a NIST-traceable calibration
  • Easily re-calibrated without opening the case, saving the user time and money
  • Heating and cooling rates can be adjusted via the front panel
  • Thermal switches can be checked for actuation testing
  • RS-232 interface for communication with the user’s PC
  • Using this interface and the supplied Fluke Interface-It Software, the user is able to adjust set points and ramp rates; record dry-well readings, saving the measurements to a file; collect data whilst performing a thermal switch test, and create an electronic strip chart
  • Compact, lightweight and portable making it suitable for applications within calibration laboratories and in-the-field
  • An efficient, comprehensive, accurate and practical temperature calibrator

What’s Included?

  • Your Choice of Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert
  • NIST-traceable calibration
  • RS-232 Interface
  • Fluke’s Interface-It Software

Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert Technical Specifications

The Fluke 9103-X-256 Below Ambient Dry Well Calibrator – Choice of Insert has a sister product, the Fluke 9140-X-256 Field Dry Well Temperature Calibrator – Choice of Disc; please find below the technical specifications for both models.

Specifications 9103 9140
Range  –25 °C to 140 °C
(–13 °F to 284 °F)
at 23 °C ambient 
35 °C to 350 °C
(95 °F to 662 °F) 
Accuracy  ± 0.25 °C (holes greater than 1/4" [6.35 mm]: ± 1 °C)  ± 0.5 °C (holes greater than 1/4" [6.35 mm]: ± 1 °C) 
Stability  ± 0.02 °C at –25 °C
± 0.04 °C at 140 °C 
± 0.03 °C at 50 °C
± 0.05 °C at 350 °C 
Well-to-Well Uniformity ± 0.1 °C between similarly sized wells ± 0.1 °C with similarly sized wells 
Heating Times  18 minutes from ambient to 140 °C  12 minutes from ambient to 350 °C 
Cooling Times  20 minutes from ambient to –25 °C  15 minutes from 350 °C to 100 °C 
Stabilization Time  7 minutes 
Immersion Depth  124 mm (4.875") 
Inserts Insert A, B, C, or D included (specify when ordering) 
Outside Insert Dimensions  31.8 mm dia. x 124 mm (1.25 x 4.88 in) 
Computer Interface  RS-232 interface
Power  115 V ac (± 10 %), 1.3 A or 230 V ac (± 10 %), 0.7 A, switchable, 50/60 Hz, 150 W  115 V ac (± 10 %), 4.4 A or 230 V ac (± 10 %), 2.2 A, switchable, 50/60 Hz, 500 W 
Size (WxHxD)  143 x 261 x 245 mm
(5.63 x 10.25 x 9.63 in) 
152 x 86 x 197 mm
(6 x 3.375 x 7.75 in) 
Weight  5.7 kg (12 lb)  2.7 kg (6 lb) 
NIST-Traceable Certificate  Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, and 140 °  Data at 50 °C, 100 °C, 150 °C, 200 °C, 250 °C, 300 °C, and 350 °C 
Fluke Industrial Temperature Calibrator Selection Guide Manuals & Guides Fluke 9103-X-256 Below Ambient Dry Well Calibrator - User Guide Supplement Manuals & Guides Fluke 9103-X-256 Below Ambient Dry Well Calibrator - User Guide Manuals & Guides

Reviews

Additional Information
SKU Fluke 9103-X-256
Manufacturer Part Code 3015846

Made to Order item: This item is made-to-order with a lead time once the order is placed. Once an order is placed it cannot be cancelled or returned unless the unit is found to be faulty.