The upgraded version of Fluke's 753, the 754 can do everything the 753 can do plus it includes HART communication capability!
Designed to support the commands in HART protocol version 5.7, the 754 includes several different HART functions including universal commands (read manufacturer and device type, read primary variable and more), common practice comments (read multiple variables, set damping time or perform loop tests) and device-specific commands (e.g. sensor trim). It also includes a point-to-point operation for connecting the 754 to a single HART device using a 4-20mA loop, a multi-drop mode for bussing together several HART instruments and a Burst mode for transmitting bursts of data from the 754 without waiting to be interrogated by a master unit.
On top of its HART functionality, the Fluke 754 is also a multifunctional process calibrator. This means it can measure, source and simulate many parameters, including RTDs, pressure and electrical signals. The information is directly logged into the instrument's built-in memory and can be downloaded and interacted with via PC.
Optional Fluke Premium Care - 1 Year (UK Only)
- Accidental damage and repair costs included
- Expedited calibration and repair
- Annual tool inspection, calibration, and updating
- Replacement of damaged accessories (Some limitations apply, see Fluke Terms and Conditions)
Select the Premium Care option in the dropdown box when choosing the model you wish to purchase, only available for users in the UK. Please visit our Blog for more information.
Fluke 754 Documenting Process Calibrator Key Features
- HART communication functionality
- Measure volts, mA, RTDs, thermocouples, frequency and ohms (ideal for testing sensors, transmitters and other instruments)
- Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms and pressure
- Power transmitters during the test using a loop supply with simultaneous mA measurements
- Compatible with 29 Fluke 700Pxx Pressure Modules (sold separately)
- Create automated calibration procedures to speed up your testing time
- Record and document results on built-in memory that can store a full week of download procedures and calibration results
- Advanced features - autostep, custom units, user-entered values during test, one-point and two-point switch testing, square root DP flow testing and programmable measurement delay
- Easy to use - minimal configuration options and advanced features make calibration a breeze
- Rechargeable batteries - 10 hours of use on a full charge
- Handling of fast-pulsed RTD transmitters and PLCs with pulses as short as 1ms
What's Included?
- Fluke 754 Documenting Process Calibrator
- 3 x Sets of Stackable Test Leads
- 3 x Sets of TP220 Test Probes with 3 x Extended Tooth Alligator Clips
- 2 x Sets of AC280 Hook Clips
- BP7240 Li-Ion Battery Pack
- BC7240 Battery Charger
- C799 Field Soft Case
- USB Communication Cable
- Getting Started Guide
- Instruction Manual on CD-ROM
- NIST Traceable Certificate of Calibration
- DPC/TRACK 2 Sample Software
- HART Communication Cable
- Optional 1-Year Premium Care (UK Only)
Product Range
- Fluke 754 Documenting Process Calibrator with HART
- Fluke 754/FPC EU Documenting Process Calibrator with HART & 1-Year Premium Care (UK Only)
Measurement Accuracy | |||||
---|---|---|---|---|---|
Voltage DC | Range/Resolution | 1 Year | 2 Years | ||
100.000 mV | 0.02% + 0.005 mV | 0.03% + 0.005 mV | |||
3.00000 V | 0.02% + 0.00005 V | 0.03% + 0.00005 V | |||
30.0000 V | 0.02% + 0.0005 V | 0.03% + 0.0005 V | |||
300.00 V | 0.05% + 0.05 V | 0.07% + 0.05 V | |||
Voltage AC | 3.000 V (40 Hz to 500 Hz) / 0.001 V | 0.5% + 0.002 V | 1.0% + 0.004 V | ||
30.00 V (40 Hz to 500 Hz) / 0.01 V | 0.5% + 0.02 V | 1.0% + 0.04 V | |||
300.0 V (40 Hz to 500 Hz) / 0.1 V | 0.5% + 0.2 V | 1.0% + 0.2 V | |||
Current DC | 30.000 mA | 0.01% + 5 uA | 0.015% + 7 uA | ||
110.00 mA | 0.01% + 20 uA | 0.015% + 30 uA | |||
Resistance | 10.000 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | ||
100.00 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | |||
1.0000 kΩ | 0.05% + 500 mΩ | 0.07% + 0.5 Ω | |||
10.000 kΩ | 0.1% + 10 Ω | 0.15% + 15 Ω | |||
Frequency | 1.00 to 110.00 Hz / 0.01 Hz | 0.05 Hz | |||
110.1 to 1100.0 Hz / 0.1 Hz | 0.5 Hz | ||||
1.101 to 11.000 kHz / 0.001 kHz | 0.005 kHz | ||||
11.01 to 50.00 kHz / 0.01 kHz | 0.05 kHz | ||||
Source Accuracy | |||||
1 Year | 2 Years | ||||
Voltage DC | 100.000 mV | 0.01% + 0.005 mV | 0.015% + 0.005 mV | ||
1.00000 V | 0.01% + 0.00005 V | 0.015% + 0.0005 V | |||
15.0000 V | 0.01% + 0.0005 V | 0.015% + 0.0005 V | |||
Current DC | 22.000 mA (source) | 0.01% + 0.003 mA | 0.02% + 0.003 mA | ||
Current sink (simulate) | 0.02% + 0.007 mA | 0.04% + 0.007 mA | |||
Resistance | 10.000 Ω | 0.01% + 10 mΩ | 0.015% + 15 mΩ | ||
100.00 Ω | 0.01% + 20 mΩ | 0.015% + 30 mΩ | |||
1.0000 kΩ | 0.02% + 0.2 Ω | 0.03% + 0.3 Ω | |||
10.000 kΩ | 0.02% + 3 Ω | 0.03% + 5 Ω | |||
Frequency | 0.1 to 10.99 Hz | 0.01 Hz | |||
0.01 to 10.99 Hz | 0.01 Hz | ||||
11.00 to 109.99 Hz | 0.1 Hz | ||||
110.0 to 1099.9 Hz | 0.1 Hz | ||||
1.100 to 21.999 kHz | 0.002 kHz | ||||
22.000 to 50.000 kHz | 0.005 kHz | ||||
Technical Data | |||||
Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | |||
Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | ||||
Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | ||||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Rate | 4 steps/second | ||||
Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | ||||
Loop power function | Voltage | Selectable, 26 V | |||
Accuracy | 10%, 18 V minimum at 22 mA | ||||
Maximum current | 25 mA, short circuit protected | ||||
Maximum input voltage | 50 V DC | ||||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Manual step | Selectable step, change with arrow buttons | ||||
Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | ||||
Environmental Specifications | |||||
Operating temperature | -10°C to +50°C | ||||
Storage temperature | -20°C to +60°C | ||||
Dust/water resistance | Meets IP52, IEC 529 | ||||
Operating altitude | 3000 m above mean sea level (9842 ft) | ||||
Safety Specifications | |||||
Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 | ||||
Mechanical and General Specifications | |||||
Size | 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) | ||||
Weight | 1.2 kg (2.7 lb) | ||||
Batteries | Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh | ||||
Battery life | >8 hours typical | ||||
Battery replacement | Replace without opening the calibrator; no tools required | ||||
Side port connections | Pressure module connector | ||||
USB Connector to interface to your PC | |||||
Digital instrument (HART) connector | |||||
Connection for optional battery charger/eliminator | |||||
Data storage capacity | 1 week of calibration procedures results | ||||
90 day specifications | The standard specification interval for the 750 Seriesare 1 and 2 years. | ||||
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. | |||||
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant. | |||||
Temperature, Resistance Temperature Detectors | |||||
Degrees or % of reading - Type (α) | Range °C | Measure °C1 | |||
1 year | 2 year | ||||
100 Ω Pt (385) | -200 to 100 100 to 800 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
200 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
500 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
1000 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
100 Ω Pt (3916) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
100 Ω Pt (3926) | -200 to 100 100 to 630 |
0.08°C 0.02% + 0.06°C |
0.16°C 0.04% + 0.12°C |
||
10 Ω Cu (427) | -100 to 260 | 0.2°C | 0.4°C | ||
120 Ω Ni (672) | -80 to 260 | 0.1°C | 0.2°C | ||
Source current | Source °C | Allowable current2 | |||
1 year | 2 year | ||||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
500 μA | 0.06°C 0.017% + 0.05°C |
0.12°C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
250 μA | 0.06°C 0.017% + 0.05°C |
0.12°C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
150 μA | 0.06 C 0.017% + 0.05°C |
0.12 C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
3 mA | 0.2°C | 0.4°C | 0.1 mA to 10 mA | ||
1 mA | 0.04°C | 0.08°C | 0.1 mA to 10 mA | ||
1. For two and three-wire RTD measurements, add 0.4°C to the specifications. 2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms |
|||||
Temperature, Thermocouples | |||||
Type | Source °C | Measure °C | Source °C | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000 | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |