All Fluke Products

Fluke 753 Documenting Process Calibrator

SKU
FLUKE753
Rating:
0%
(0)
Reward Points
:
5700
£6,840.00 £5,700.00
Free UK Delivery*
Advice & Support Call 01642 931329 or contact us
Price Match We won't be beaten, let us know
Fluke 753 Documenting Process Calibrator

Powerful, multifunctional and condensed down into a portable instrument, the Fluke 753 is a calibrator that can measure and source everything from temperature to pressure. It can be used in practically any calibration scenario and gives you all of the tools you'll need in one comprehensive instrument.

The 753 is truly a multifunction instrument; it can measure voltage, mA, RTDs, thermocouples, frequency and ohms, source/simulate voltage, mA, RTDs, frequency, ohms and pressure and includes dedicated memory for storing up to a full week of downloaded procedures and calibration results.

With the 753 at hand, you can configure automated calibration procedures for linear transmitters, DP flow transmitters and one or two-point limit switches. This means you only have to select your configuration, connect up and start testing with absolutely minimal effort on your part.

When measuring pressure, you can use Fluke's expansive range of pressure modules (sold separately) in combination with this instrument to cater for practically any pressure application ranging from gauge to vacuum.

Fluke 753 Documenting Process Calibrator Key Features

  • Measure volts, mA, RTDs, thermocouples, frequency and ohms (ideal for testing sensors, transmitters and other instruments)
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms and pressure
  • Power transmitters during test using loop supply with simultaneous mA measurements
  • Compatible with 29 Fluke 700Pxx Pressure Modules (sold separately)
  • Create automated calibration procedures to speed up your testing time
  • Record and document results on built-in memory that can store a full week of download procedures and calibration results
  • Advanced features - autostep, custom units, user-entered values during test, one-point and two-point switch testing, square root DP flow testing and programmable measurement delay
  • Easy to use - minimal configuration options and advanced features make calibration a breeze
  • Rechargeable batteries - 10 hours of use on a full charge
  • Handling of fast-pulsed RTD transmitters and PLCs with pulses as short as 1ms

What's Included?

  • Fluke 753 Documenting Process Calibrator
  • 3 x Sets of Stackable Test Leads
  • 3 x Sets of TP220 Test Probes with 3 x Extended Tooth Alligator Clips
  • 2 x Sets of AC280 Hook Clips
  • BP7240 Li-Ion Battery Pack
  • BC7240 Battery Charger
  • C799 Field Soft Case
  • USB Communication Cable
  • Getting Started Guide
  • Instruction Manual on CD-ROM
  • NIST Traceable Certificate of Calibration
  • DPC/TRACK 2 Sample Software

Fluke 735 Technical Specifications

Measurement Accuracy
Voltage DC Range /Resolution 1 Year 2 Years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Current DC 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω
10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω
Frequency 1.00 to 110.00 Hz / 0.01 Hz
0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz
0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz
0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz
0.05 kHz
Source Accuracy


1 Year 2 Years
Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ
100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
Frequency 0.1 to 10.99 Hz
0.01 Hz
0.01 to 10.99 Hz
0.01 Hz
11.00 to 109.99 Hz
0.1 Hz
110.0 to 1099.9 Hz
0.1 Hz
1.100 to 21.999 kHz
0.002 kHz
22.000 to 50.000 kHz
0.005 kHz
Technical Data
Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Rate 4 steps/second
Trip detect Continuity or voltage (continuity detection not available when sourcing current)
Loop power function Voltage Selectable, 26 V
Accuracy 10%, 18 V minimum at 22 mA
Maximum current 25 mA, short circuit protected
Maximum input voltage 50 V DC
Step functions Source functions Voltage, current, resistance, frequency, temperature
Manual step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature -10°C to +50°C
Storage temperature -20°C to +60°C
Dust/water resistance Meets IP52, IEC 529
Operating altitude 3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight 1.2 kg (2.7 lb)
Batteries Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh
Battery life > 8 hours typical
Battery replacement Via snap-shut door without opening calibrator; no tools required
Side port connections Pressure module connector
USB Connector to interface to your PC
Connection for optional battery charger/eliminator
Data storage capacity 1 week of calibration procedures results
90 day specifications The standard specification interval for the 750 Series are 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Temperature, Resistance Temperature Detectors
Degrees or % of reading - Type (α) Range °C Measure °C1
  1 year 2 year
100 Ω Pt (385) -200 to 100
100 to 800
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
200 Ω Pt (385) -200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
500 Ω Pt (385) -200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
1000 Ω Pt (385) -200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3916) -200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3926) -200 to 100
100 to 630
0.08°C
0.02% + 0.06°C
0.16°C
0.04% + 0.12°C
10 Ω Cu (427) -100 to 260 0.2°C 0.4°C
120 Ω Ni (672) -80 to 260 0.1°C 0.2°C
Source current Source °C Allowable current2
  1 year 2 year  
1 mA 0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
500 μA 0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
250 μA 0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
150 μA 0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
1 mA 0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
1 mA 0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
3 mA 0.2°C 0.4°C 0.1 mA to 10 mA
1 mA 0.04°C 0.08°C 0.1 mA to 10 mA
1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples
Type Source °C Measure °C Source °C


1 year 2 years 1 year 2 years
E -250 to -200 1.3 2.0 0.6 0.9
-200 to -100 0.5 0.8 0.3 0.4
-100 to 600 0.3 0.4 0.3 0.4
600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
-100 to 900 0.5 0.8 0.5 0.8
900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1.0 0.4 0.6
-100 to 400 0.3 0.4 0.3 0.4
400 to 1200 0.5 0.8 0.3 0.4
1200 to 1372 0.7 1.0 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
-200 to 0 0.6 0.9 0.4 0.6
0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2.0 1.0 1.5
800 to 1000 1.0 1.5 0.8 1.2
1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
100 to 1767 1.0 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
200 to 1400 0.9 1.4 0.9 1.4
1400 to 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
800 to 1200 0.8 1.2 0.7 1.0
1200 to 1800 1.1 1.6 0.9 1.4
1800 to 2316 2.0 3.0 1.3 2.0
L -200 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
1000 to 2000 1.6 2.4 0.6 0.9
2000 to 2500 2.0 3.0 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6
Accessories related to this product:
Fluke 750 Series Documenting Process Calibrators - Datasheet Datasheets Fluke 750 Series Documenting Process Calibrators - User Manual Manuals & Guides

Reviews

Additional Information
SKU FLUKE753
Manufacturer Part Code 3977641
Country of Origin United States