Powerful, multifunctional and condensed down into a portable instrument, the Fluke 753 is a calibrator that can measure and source everything from temperature to pressure. It can be used in practically any calibration scenario and gives you all of the tools you'll need in one comprehensive instrument.
The 753 is truly a multifunction instrument; it can measure voltage, mA, RTDs, thermocouples, frequency and ohms, source/simulate voltage, mA, RTDs, frequency, ohms and pressure and includes dedicated memory for storing up to a full week of downloaded procedures and calibration results.
With the 753 at hand, you can configure automated calibration procedures for linear transmitters, DP flow transmitters and one or two-point limit switches. This means you only have to select your configuration, connect up and start testing with absolutely minimal effort on your part.
When measuring pressure, you can use Fluke's expansive range of pressure modules (sold separately) in combination with this instrument to cater for practically any pressure application ranging from gauge to vacuum.
Fluke 753 Documenting Process Calibrator Key Features
- Measure volts, mA, RTDs, thermocouples, frequency and ohms (ideal for testing sensors, transmitters and other instruments)
- Source/simulate volts, mA, thermocouples, RTDs, frequency, ohms and pressure
- Power transmitters during test using loop supply with simultaneous mA measurements
- Compatible with 29 Fluke 700Pxx Pressure Modules (sold separately)
- Create automated calibration procedures to speed up your testing time
- Record and document results on built-in memory that can store a full week of download procedures and calibration results
- Advanced features - autostep, custom units, user-entered values during test, one-point and two-point switch testing, square root DP flow testing and programmable measurement delay
- Easy to use - minimal configuration options and advanced features make calibration a breeze
- Rechargeable batteries - 10 hours of use on a full charge
- Handling of fast-pulsed RTD transmitters and PLCs with pulses as short as 1ms
What's Included?
- Fluke 753 Documenting Process Calibrator
- 3 x Sets of Stackable Test Leads
- 3 x Sets of TP220 Test Probes with 3 x Extended Tooth Alligator Clips
- 2 x Sets of AC280 Hook Clips
- BP7240 Li-Ion Battery Pack
- BC7240 Battery Charger
- C799 Field Soft Case
- USB Communication Cable
- Getting Started Guide
- Instruction Manual on CD-ROM
- NIST Traceable Certificate of Calibration
- DPC/TRACK 2 Sample Software
Fluke 735 Technical Specifications
Measurement Accuracy | |||||
---|---|---|---|---|---|
Voltage DC | Range /Resolution | 1 Year | 2 Years | ||
100.000 mV | 0.02% + 0.005 mV | 0.03% + 0.005 mV | |||
3.00000 V | 0.02% + 0.00005 V | 0.03% + 0.00005 V | |||
30.0000 V | 0.02% + 0.0005 V | 0.03% + 0.0005 V | |||
300.00 V | 0.05% + 0.05 V | 0.07% + 0.05 V | |||
Voltage AC | 3.000 V (40 Hz to 500 Hz) / 0.001 V | 0.5% + 0.002 V | 1.0% + 0.004 V | ||
30.00 V (40 Hz to 500 Hz) / 0.01 V | 0.5% + 0.02 V | 1.0% + 0.04 V | |||
300.0 V (40 Hz to 500 Hz) / 0.1 V | 0.5% + 0.2 V | 1.0% + 0.2 V | |||
Current DC | 30.000 mA | 0.01% + 5 uA | 0.015% + 7 uA | ||
110.00 mA | 0.01% + 20 uA | 0.015% + 30 uA | |||
Resistance | 10.000 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | ||
100.00 Ω | 0.05% + 50 mΩ | 0.07% + 70 mΩ | |||
1.0000 kΩ | 0.05% + 500 mΩ | 0.07% + 0.5 Ω | |||
10.000 kΩ | 0.1% + 10 Ω | 0.15% + 15 Ω | |||
Frequency | 1.00 to 110.00 Hz / 0.01 Hz | 0.05 Hz | |||
110.1 to 1100.0 Hz / 0.1 Hz | 0.5 Hz | ||||
1.101 to 11.000 kHz / 0.001 kHz | 0.005 kHz | ||||
11.01 to 50.00 kHz / 0.01 kHz | 0.05 kHz | ||||
Source Accuracy | |||||
1 Year | 2 Years | ||||
Voltage DC | 100.000 mV | 0.01% + 0.005 mV | 0.015% + 0.005 mV | ||
1.00000 V | 0.01% + 0.00005 V | 0.015% + 0.0005 V | |||
15.0000 V | 0.01% + 0.0005 V | 0.015% + 0.0005 V | |||
Current DC | 22.000 mA (source) | 0.01% + 0.003 mA | 0.02% + 0.003 mA | ||
Current sink (simulate) | 0.02% + 0.007 mA | 0.04% + 0.007 mA | |||
Resistance | 10.000 Ω | 0.01% + 10 mΩ | 0.015% + 15 mΩ | ||
100.00 Ω | 0.01% + 20 mΩ | 0.015% + 30 mΩ | |||
1.0000 kΩ | 0.02% + 0.2 Ω | 0.03% + 0.3 Ω | |||
10.000 kΩ | 0.02% + 3 Ω | 0.03% + 5 Ω | |||
Frequency | 0.1 to 10.99 Hz | 0.01 Hz | |||
0.01 to 10.99 Hz | 0.01 Hz | ||||
11.00 to 109.99 Hz | 0.1 Hz | ||||
110.0 to 1099.9 Hz | 0.1 Hz | ||||
1.100 to 21.999 kHz | 0.002 kHz | ||||
22.000 to 50.000 kHz | 0.005 kHz | ||||
Technical Data | |||||
Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | |||
Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | ||||
Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | ||||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Rate | 4 steps/second | ||||
Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | ||||
Loop power function | Voltage | Selectable, 26 V | |||
Accuracy | 10%, 18 V minimum at 22 mA | ||||
Maximum current | 25 mA, short circuit protected | ||||
Maximum input voltage | 50 V DC | ||||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Manual step | Selectable step, change with arrow buttons | ||||
Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | ||||
Environmental Specifications | |||||
Operating temperature | -10°C to +50°C | ||||
Storage temperature | -20°C to +60°C | ||||
Dust/water resistance | Meets IP52, IEC 529 | ||||
Operating altitude | 3000 m above mean sea level (9842 ft) | ||||
Safety Specifications | |||||
Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 | ||||
Mechanical and General Specifications | |||||
Size | 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) | ||||
Weight | 1.2 kg (2.7 lb) | ||||
Batteries | Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh | ||||
Battery life | > 8 hours typical | ||||
Battery replacement | Via snap-shut door without opening calibrator; no tools required | ||||
Side port connections | Pressure module connector | ||||
USB Connector to interface to your PC | |||||
Connection for optional battery charger/eliminator | |||||
Data storage capacity | 1 week of calibration procedures results | ||||
90 day specifications | The standard specification interval for the 750 Series are 1 and 2 years. | ||||
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. | |||||
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant. | |||||
Temperature, Resistance Temperature Detectors | |||||
Degrees or % of reading - Type (α) | Range °C | Measure °C1 | |||
1 year | 2 year | ||||
100 Ω Pt (385) | -200 to 100 100 to 800 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
200 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
500 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
1000 Ω Pt (385) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
100 Ω Pt (3916) | -200 to 100 100 to 630 |
0.07°C 0.02% + 0.05°C |
0.14°C 0.04% + 0.10°C |
||
100 Ω Pt (3926) | -200 to 100 100 to 630 |
0.08°C 0.02% + 0.06°C |
0.16°C 0.04% + 0.12°C |
||
10 Ω Cu (427) | -100 to 260 | 0.2°C | 0.4°C | ||
120 Ω Ni (672) | -80 to 260 | 0.1°C | 0.2°C | ||
Source current | Source °C | Allowable current2 | |||
1 year | 2 year | ||||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
500 μA | 0.06°C 0.017% + 0.05°C |
0.12°C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
250 μA | 0.06°C 0.017% + 0.05°C |
0.12°C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
150 μA | 0.06°C 0.017% + 0.05°C |
0.12°C 0.034% + 0.10°C |
0.1 mA to 1 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
1 mA | 0.05°C 0.0125% + 0.04°C |
0.10°C 0.025% + 0.08°C |
0.1 mA to 10 mA | ||
3 mA | 0.2°C | 0.4°C | 0.1 mA to 10 mA | ||
1 mA | 0.04°C | 0.08°C | 0.1 mA to 10 mA | ||
1. For two and three-wire RTD measurements, add 0.4°C to the specifications. 2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms |
|||||
Temperature, Thermocouples | |||||
Type | Source °C | Measure °C | Source °C | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000 | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |