The Fluke 7526A Precision Process Calibrator can be used to calibrate the following components:
- RTD and thermocouple readouts
- Pressure gauges
- Temperature transmitters
- Digital process simulators
- Data loggers
- Multimeters
The Fluke 7526A Precision Process Calibrator is able to calibrate the above modules and instruments as a result of its extensive capabilities. It is able to simulate and measure nine types of RTDs and thirteen types of thermocouples. In addition to this, it can source between 0 and 100V of DC voltage and is able to measure DC voltage between 0 and 100V to an accuracy of 0.004% of the reading. Furthermore, this instrument can source 0 to 100mA of DC current and can measure DC current between 0 and 50mA to an accuracy of 0.01%. Adding to its sourcing capabilities is its ability to issue a 24V DC loop power supply, whilst its ability to measure 4-20mA loop current extends its measurement capabilities.
As well as the above, the Fluke 7526A Precision Process Calibrator can be used with Fluke’s 700 or 525A-P Pressure Modules in order to measure pressure to within 0.008% of the determined pressure value. The Fluke 7526A Precision Process Calibrator also accepts ITS-90 coefficients allowing it to take accurate SPRT (Standard Platinum Resistance Thermometers) measurements.
This instrument is also compatible with Fluke's MET/CAL® Software (NOT included) which allows the user to monitor measurements and reports, as well as conduct automated calibration procedures.
In addition to conducting automated calibration procedures using the MET/CAL® Software, the Fluke 7526A Precision Process Calibrator also includes an automated switch-test function.
In order to facilitate all the Fluke 7526A Precision Process Calibrator’s measurement capabilities, it includes a number of physical features, including:
- Isolated pressure module input: automatically detects the type and value of the Fluke 700 Pressure Module connected
- DC voltage, current and switch-test input and 24V DC power supply
- Thermocouple input and output terminal
- Two-wire RTD/Ω output terminal
- Four-wire RTD/Ω input terminals
- Isolated input controls: HART key activates a 250Ω loop resistor facilitating the HART device's ability to read a digital HART signal superimposed onto the 4-20mA loop current
- Primary input and output controls
- Cursor controls
This conveniently designed pressure calibrator is ideal for the benchtop calibration of a variety of temperature and process pressure equipment. The Fluke 7526A Precision Process Calibrator is a versatile, accurate, ergonomic and cost-effective instrument.
Fluke 7526A Precision Process Calibrator Key Features
- Ideal for calibrating a wide variety of temperature and pressure equipment
- Suitable for calibrating RTDs, thermocouple readouts, pressure gauges, temperature transmitters, digital process simulators, data loggers and multimeters
- Simulate and measure 9 types of RTDs
- Simulate and measure 13 types of thermocouples
- Source and measure 0 to 100V DC
- Measure DC voltage to an accuracy of 0.004% of the reading
- Source 0mA to 100mA DC current
- Measure DC current between 0 and 50mA to an accuracy of 0.01%
- Issue 24VDC loop power supply
- Measure 4-20mA loop current
- Use with Fluke’s 700 or 525A-P Pressure Modules to measure pressure to an accuracy of 0.008% of the reading
- Accepts ITS-90 coefficients for accurate SPRT measurements
- Compatible with Fluke’s MET/CAL® Software (NOT included)
- Automated switch-test function
- Isolated pressure module input: automatically detects the type and value of the Fluke 700 Pressure Module connected
- DC voltage, current and switch-test input and 24V DC power supply
- Thermocouple input and output terminal
- Two-wire RTD/Ω output terminal
- Four-wire RTD/Ω input terminals
- Isolated input controls: HART key activates a 250Ω loop resistor facilitating the HART device's ability to read a digital HART signal superimposed onto the 4-20mA loop current
- Primary input and output controls
- Cursor controls
- Ideal for benchtop calibration
- Versatile, accurate, ergonomic and cost-effective
What’s Included?
- Fluke 7526A Precision Process Calibrator
- Traceable Calibration Report
- User Manual CD
- Getting Started Guide
- Power Cord
- Thermocouple Shorting Jumper
- USB-to-Serial Adapter Cable
Fluke 7526A Precision Process Calibrator Technical Specifications
Temperature Range (°C)
Absolute Uncertainty,
tcal ±5 °C, ± (°C), 1-yr
DC Voltage Output | |||
Range [1] | Absolute Uncertainty, ± (ppm of output + μV), 1-yr |
Resolution | |
0 to 100 mV | 30 | 3 | 1 μV |
0 to 1 V | 30 | 10 | 10 μV |
0 to 10 V | 30 | 100 | 100 μV |
0 to 100 V | 30 | 1 mV | 1 mV |
TC Output and Input | |||
–10 mV to 75 mV | 30 | 2 μV | 1 μV |
[1] All outputs are positive only unless otherwise noted | |||
DC voltage, isolated input | |||
Range | Absolute Uncertainty, ± (ppm of reading + mV), 1-yr |
Resolution | |
0 to 10 V | 50 | 0.2 | 100 μV |
10 V to 100 V | 50 | 2.0 | 1 mV |
DC Current, output | |||
Range[1] | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr |
Resolution | |
0 to 100 mA | 50 | 1 | 1 μA |
[1] For line voltages less than 95 V, ±100 ppm of reading | |||
DC current, isolated input | |||
Range | Absolute Uncertainty, ± (ppm of reading + μA), 1-yr |
Resolution | |
0 mA to 50 mA | 100 | 1 | 0.1 μA |
0 mA to 24 mA [1][2] (Loop Power) |
100 | 1 | 0.1 μA |
[1] Loop Power: 24 V ±10% [2] HART Resistor: 250 Ω ± 3% |
|||
Resistance, output | |||
Range | Absolute Uncertainty, ± (ppm of reading + Ω), 1-yr |
Resolution | |
0 Ω to 400 Ω | 20 | 0.004 | 0.001 Ω |
0 Ω to 4 kΩ | 20 | 0.04 | 0.01 Ω |
Sample thermocouple accuracy, input/output (does not include all available TC types)[1] |
|||
TC type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr [2] |
|
Min | Max | ||
J | -210 | 1200 | 0.09 |
K | -250 | 1372 | 0.1 |
S | -50 | 1767 | 0.29 |
T | -250 | 400 | 0.11 |
[1] See extended specifications for all TC types (B,C,E,J,K,L,N,R,S,T,U,XK,BP) [2] Best accuracy within specified TC temperature range |
|||
Sample RTD and thermistor, output (does not include all available RTD types)[1] |
|||
RTD Type | Temperature Range (°C) | Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr |
|
Min | Max | ||
Pt 385, 100 Ω | –80 | 100 | 0.020 |
100 | 300 | 0.024 | |
300 | 800 | 0.038 | |
YSI 400 | 15 | 50 | 0.007 |
[1] See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt- 200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT | |||
General Specifications | |||
Standard interface | RS-232, IEEE-488 (GPIB) | ||
Temperature performance |
Operating: | 0 °C to 50 °C | |
Calibration (tcal): | 18 °C to 28 °C | ||
Storage: | –20 °C to 70 °C | ||
Electromagnetic compatibility |
CE: Conforms to EN61326; operation in controlled EM environments |
||
Temperature coefficient | Temperature coefficient for temperatures outside tcal 5 °C is 10% of the 90-day specification (or 1 year if applicable) per °C |
||
Relative humidity | Operating: | <80% to 30 °C | |
<70% to 40 °C | |||
<40% to 50 °C | |||
Altitude | Operating: | 3,000 m (9,800 ft) max | |
Non-operating: | 12,200 m (40,000 ft) max | ||
Safety | EN/IEC 61010-1:2010 3rd Edition, UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12 |
||
Analog low isolation | 20 V | ||
Line power | 120 V~: | 100 V to 120 V | |
240 V~: | 220 V to 240 V | ||
Line frequency | 47 Hz to 63 Hz | ||
Line voltage variation | ± 10 % about setting | ||
Power consumption | 15 VA maximum | ||
Dimensions | Height: | 14.6 cm (5.75 in) | |
Width: | 44.5 cm (17.5 in) | ||
Depth: | 29.8 cm (11.75 in) | ||
Weight (without options) |
4.24 kg (9.35 lb) |