Gage pressure modules have one pressure fitting and measure the pressure concerning atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. Each module is clearly labelled for range, overpressure, and media compatibility. All modules include NPT, metric (BSP), and M20 adapters. Available from 0 to 15 Psi (0 to 1 bar) to 0 to 10000 Psi (0 to 700 bar).
Reference Module Specifications | ||||
---|---|---|---|---|
Model | Parameter / Range | Burst Rating5 | Hi Side Media2 | Lo Side Media2 |
750R044 | 0 to 15 psi (0 to 1 bar) | 3X | Dry air | Dry air |
750R064 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | — |
750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | — |
750R084 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | — |
750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | — |
750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | — |
750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | — |
750R314 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | — |
Model | Reference Uncertainty3 | Total Uncertainty 1-Year (15-35°C) | Total Uncertainty 1-Year1 | Total Uncertainty 6-Month (15-35°C) | Total Uncertainty 6-Month1 |
---|---|---|---|---|---|
750R044 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R064 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R07 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R084 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R27 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R29 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R30 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
750R314 | ±0.01% of FS | ±0.02% of FS | ±0.04% of FS | ±0.015% of FS | ±0.035% of FS |
1 - Total uncertainty, % of full span for temperature range 0°C to +50°C, one year interval. Total uncertainty, 1.0% of full span for temperature range -10°C to 0°C, one-year interval. No 6-month specification available for range -10°C to 0°C. 2 - “Non-corrosive gasses” indicates Dry air or non-corrosive gas as compatible media. “Stainless Steel 316-SS” indicates media compatible with Type 316 Stainless Steel. 3 - Reference Uncertainty is the specification for as left data for 24 hours. 4 - When reference class modules are used with fixed resolution products (717, 718, 719 series, 725, and 726) calibrators add ±1 count to the overall accuracy specification. 5 - Burst rating specification refers to the multiplier times the full scale of the module for the rated burst pressure. |
---|