The Fluke 9132-256 Portable IR Calibrator can be used to calibrate above ambient, handheld pyrometers, including non-contact infrared thermometers. To this end, it has a temperature range of 50 to 500°C.
The Fluke 9132-256 Portable IR Calibrator has an accuracy of ±0.5°C to ±0.8°C, depending on the temperature generated. This ensures that infrared thermometers can be calibrated accurately, as does the inclusion of a 57mm-diameter blackbody target. This target has an emissivity of 0.95 (±0.02 %) and can be accurately calibrated via the RTD-reference well located behind the target. Its large target area ensures that the field-of-view of a majority of non-contact infrared thermometers are accommodated for, thereby increasing this instrument’s scope-for-use.
The temperature of the blackbody target can be precisely controlled and may be increased or decreased in set-point increments of 0.1°C. The Fluke 9132-256 Portable IR Calibrator is able to offer a stability of ±0.1°C and takes only a short while to heat or cool to the desired temperature:
In addition to its accuracy and efficiency, this compact, ergonomic and portable instrument with a hand strap is also easy-to-use:
1.) Set the desired temperature via the front panel control buttons
2.) Wait for the Fluke 9132-256 Portable IR Calibrator to heat/cool to the desired temperature and stabilise
3.) Point the non-contact instrument-under-investigation at the blackbody target
4.) Compare the reading on the pyrometer to the reading on the Fluke 9132-256 Portable IR Calibrator’s display
5.) Record any difference in the readings
This instrument can be adjusted to suit the user’s needs and to this end includes an eight set-point memory and option to display readings in °C or °F. Its RS-232 interface facilitates the user’s ability to configure the instrument using his/her PC and the included Fluke Model 9930 Interface-It Software.
The Fluke 9132-256 Portable IR Calibrator is a simple-to-use, practical, accurate and reliable temperature calibrator, ideal for checking a variety of pyrometers, especially non-contact infrared thermometers.
Spend £150+ (Ex VAT) on any qualified Fluke or Beha-Amprobe product(s) and claim a free tool of your choice from the list at Fluke Promotion. Valid February 17 2025 - July 31 2025.
Claims are made direct to Fluke. For more information and terms and conditions, visit Fluke Promotion.
The Fluke 9132-256 Portable IR Calibrator is part of Fluke's series of 913X Portable IR Calibrators; please find below the technical specifications for both the models included in this series.
Specifications | 9133 | 9132 |
Temperature Range | –30 °C to 150 °C at 23 °C ambient (–22 °F to 302 °F at 73 °F ambient) |
50 °C to 500 °C (122°F to 932°F) |
Accuracy | ± 0.4 °C (± 0.72 °F) | ± 0.5 °C at 100 °C (± 0.9°F at 212°F) |
± 0.8 °C at 500 °C (± 1.4 °F at 932 °F) |
||
Stability | ± 0.1 °C (± 0.18 °F) | ± 0.1°C at 100°C (± 0.18°F at 212°F) |
± 0.3°C at 500°C (± 0.54°F at 932°F) |
||
Target Size | 57 mm (2.25 in) | |
Target Emissivity | 0.95 (± 0.02 from 8 to 14 µm) | |
Resolution | 0.1 ° | |
Heating Time | 15 minutes (25 °C to 150 °C) | 30 minutes (50 °C to 500 °C) |
Cooling Time | 15 minutes (25 °C to –20 °C) | 30 minutes (500 °C to 100 °C) |
Computer Interface | RS-232 interface | |
Power | 115 V ac (± 10 %), 1.5 A, or 230 V ac (± 10 %), 1.0 A, switchable, 50/60 Hz, 200 W |
115 V ac (± 10 %), 3 A or 230 V ac (± 10 %), 1.5 A, switchable, 50/60 Hz, 340 W |
Size (HxWxD) | 152 x 286 x 267 mm (6 x 11.25 x 10.5 in) |
102 x 152 x 178 mm (4 x 6 x 7 in ) |
Weight | 4.6 kg (10 lb) | 1.8 kg (4 lb) |
NIST-Traceable Contact Calibration |
Data at –30°C, 0°C, 25°C, 75°C, 100°C, 125°C, and 150°C | Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C |