The Fluke 1523-256 1-Channel Handheld Thermometer Kit 4 includes:
- 1x Fluke 1523-256 1-Channel Handheld Thermometer and Accessories
- 1x Fluke 5615-12-P Secondary Reference Temperature Probe
- 1x Fluke 2373-LTC Lemo to Universal Thermocouple Adapter
- 1x Carry Case
For further information regarding these products, please see the descriptions below.
Fluke 1523-256 1-Channel Handheld Thermometer Description
The Fluke 1523-256 1-Channel Handheld Thermometer includes three sensors which allow the user to measure, graph and record PRTs, thermistors and thermocouples. These sensors have the following measurement ranges and accuracies:
- PRT sensor: -200°C to 1000°C (±0.011°C)
- Thermocouple sensor: -200°C to 2315°C (±0.24°C for J, K, L and M)
- Thermistor sensor: -50°C to 150°C (±0.002°C)
This instrument is able to accommodate PRTs/RTDs, thermocouples and thermistors, allowing the user to adapt the Fluke 1523-256 1-Channel Handheld Thermometer to the application. It is possible to lock probes to particular channels in order to maintain system calibration traceability. The Fluke 1523-256 1-Channel Handheld Thermometer is able to identify the probe that has been connected and will recall any calibration information applicable to the connected probe from the internal INFO-CON memory chip.
Probe configurations can be created and uploaded to the Fluke 1523-256 1-Channel Handheld Thermometer using the RS-232 interface and supplied Fluke 9940 Toolkit Software. The user is also able to send commands directly to the Fluke 1523-256 1-Channel Handheld Thermometer using the RS-232 interface and Toolkit software; additionally, he/she may log and organise data recorded by the Fluke 1523-256 1-Channel Handheld Thermometer using these accessories.
Alternatively, up to twenty-five measurements with statistics can be saved directly to the Fluke 1523-256 1-Channel Handheld Thermometer. As standard the Fluke 1523-256 1-Channel Handheld Thermometer has a sample rate of one reading per second, however, when used in Fast Mode this is increased to one reading per 0.3 seconds.
Measurements are viewable via the 128 x 64-pixel, backlit, LCD display. Graphs are plotted and scaled in real-time and are also viewable via this display. Using the dedicated function button, the user can adjust the resolution of the graph to monitor processes over time and/or determine the stability of temperatures quickly and without the use of statistics.
In order to ensure the accuracy of measurements, the Fluke 1523-256 utilises a current reversal technique in order to eliminate thermal EMFs. In addition to this, the Fluke 1523-256 1-Channel Handheld Thermometer includes precision resistors and a highly stable reference voltage source which work to maintain the Fluke 1523-256 1-Channel Handheld Thermometer’s accuracy in all external temperatures. Indeed the Fluke 1523-256 1-Channel Handheld Thermometer has been thoroughly tested in extreme temperatures and harsh vibrations in order to ensure that it remains both functional and accurate in challenging environments.
Adding to this instrument’s practicality is the inclusion of power-saving settings which may be engaged in order to extend the thermometer’s operating time. Alternatively, the operator can use the Fluke 1523-256 1-Channel Handheld Thermometer with the included DC adapter in order to extend this period of functionality. This tester requires three AA batteries to operate and is able to remain functional for up to 20 hours using this power source.
The Fluke 1523-256 1-Channel Handheld Thermometer is a practical, adaptable, versatile instrument capable of taking accurate temperature measurements of PRTs, thermistors and thermocouples. It is ideal for serving as a temperature reference in bath calibrators, dry-block calibrators, thermowells, clean rooms, engines, heat exchangers, furnaces and freezers. It is particularly well suited to calibration, loop checking, plant start-up, troubleshooting, maintenance and repair applications.
Fluke 1523-256 1-Channel Handheld Thermometer Key Features
- Can measure, graph and record PRTs, thermistors and thermocouples
- PRT sensor: -200°C to 1000°C (±0.011°C)
- Thermocouple sensor: -200°C to 2315°C (±0.24°C for J, K, L and M)
- Thermistor sensor: -50°C to 150°C (±0.002°C)
- Can accommodate PRTs/RTDs, thermocouples and thermistors
- Lock probes to particular channels in order to maintain system calibration traceability
- Will identify the probe that is connected
- Will recall the relevant calibration information from the INFO-CON memory chip
- Probe configurations can be created and uploaded to the thermometer using the RS-232 interface and Fluke 9940 Toolkit Software
- Send commands directly to the thermometer using this software and interface
- Log and organise data using the interface and software
- Save up to 25 measurements with statistics to the thermometer
- A standard sample rate of one reading per second
- A Fast Mode sample rate of one reading per 0.3 seconds
- 128 x 64-pixel, backlit, LCD display
- Graphs are plotted and scaled in real-time
- Measurements and graphs can be viewed via the display
- Function button allows the user to adjust the resolution of the display
- Monitor processes over time and determine temperature stability quickly and without statistics
- Utilises a current reversal technique in order to eliminate thermal EMFs
- Includes precision resistors
- Includes a highly stable reference voltage source
- Tested in extreme temperatures and harsh vibrations
- Power-saving settings
- Can be used with the included DC adapter to extend the thermometer’s period of functionality
- Powered by three AA batteries
- Three AA batteries provide the instrument with up to 20 hours of power
- Practical, versatile, adaptable instrument
- Ideal for serving as a temperature reference
- Suitable for calibration, loop-checking, plant start-up, troubleshooting, maintenance and repair applications
What’s Included with the Fluke 1523-256 1-Channel Handheld Thermometer?
- Fluke 1523-256 1-Channel Handheld Thermometer
- NIST-traceable Certificate of Calibration
- User Guide
- CD-ROM Containing the Technical Manual
- 12V Dc Universal Power Supply
- RS-232 Cable
- Fluke 9940 I/O Toolkit Software
Fluke 5615-12-P Secondary Reference Temperature Probe Description
The Fluke 5615-12-P Secondary Reference Temperature Probe has a temperature range of -200°C to 420°C. It has an accuracy of ±0.012°C (at 0°C) and is supplied with an NVLAP accredited calibration (lab code 200706-0).
It has been calibrated from -196°C to 420°C and using ITS-90 coefficients, ensuring that it is accurate enough for use with Fluke’s more sophisticated temperature calibrators.
Further to this, the Fluke 5615-12-P Secondary Reference Temperature Probe is supplied with information regarding sensor resistance; this data is specified in 1°C increments.
In terms of its physical design, the Fluke 5615-12-P Secondary Reference Temperature Probe includes an element made of a reference-grade platinum wire (99.999% purity), ensuring that the probe remains stable. Hysteresis is all but eliminated due to the platinum wire, wound in a coil, having been placed in a mandrel, ensuring that it is uniformly supported. Furthermore, the Fluke 5615-12-P Secondary Reference Temperature Probe’s electrical configuration eradicates the effects of lead-wire resistance as it is a four-wire current-potential hookup.
This platinum resistance thermometer features an Inconel™ 600 sheath which is 305mm (12-inches) in length and 6.35mm (0.25-inches) in diameter. It is also protected by a ceramic case with a hermetic glass seal. This protects it from moisture and contaminants which thereby improves both its longevity and outward stability.
The Fluke 5615-12-P Secondary Reference Temperature Probe is a cost-effective, accurate and reliable temperature probe. It has been optimised to bridge the gap between 100Ω industrial RTDs and SPRTs, making it ideal for extending the Fluke 1523-256 1-Channel Handheld Thermometer’s capabilities.
Fluke 5615-12-P Secondary Reference Temperature Probe Key Features
- Temperature range of -200°C to 420°C
- Supplied with an NVLAP accredited calibration (lab code 200706-0)
- Calibrated from -196°C to 420°C
- Supplied with ITS-90 coefficients, making it accurate enough for use with Fluke’s more sophisticated temperature calibrators
- Supplied with information regarding sensor resistance, specified in 1°C increments
- Element made of a reference-grade platinum wire (99.999% purity)
- Reference-grade platinum ensures that the probe remains stable
- Platinum wire is wound in a coil and placed in a mandrel, ensuring that it is uniformly supported and hysteresis is all but eliminated
- The four-wire, current-potential hookup electrical configuration eradicates the effects of lead-wire resistance
- Inconel™ 600 sheath which is 305mm in length and 6.35mm in diameter
- Ceramic case with a hermetic glass seal
- The case protects the probe from moisture and contaminants, improving its longevity and stability
- A cost-effective, reliable and accurate temperature probe
- Optimised to bridge the gap between 100Ω industrial RTDs and SPRTs
- Ideal for extending the measurement capabilities of the Fluke 1523-256 1-Channel Handheld Thermometer
Fluke 2373-LTC Lemo to Universal Thermocouple Adapter Description
The Fluke 2373-LTC Lemo to Universal Thermocouple Adapter is a practical addition to the Fluke 1523-256 1-Channel Handheld Thermometer as it facilitates a connection between this instrument and thermocouples with mini-thermocouple jacks.
It includes a precision thermistor and supports reference junction compensation.
The Fluke 2373-LTC Lemo to Universal Thermocouple Adapter is ideal for expanding the Fluke 1523-256 1-Channel Handheld Thermometer’s scope-for-use.
Fluke 2373-LTC Lemo to Universal Thermocouple Adapter Key Features
- A practical addition to the Fluke 1523-256 1-Channel Handheld Thermometer
- Facilitates a connection between this instrument and thermocouples with mini-thermocouple jacks
- Includes a precision thermistor
- Supports reference junction compensation
- Ideal for expanding the Fluke 1523-256 1-Channel Handheld Thermometer’s scope-for-use
What’s Included in the Fluke 1523-256 1-Channel Handheld Thermometer Kit 4?
- 1x Fluke 1523-256 1-Channel Handheld Thermometer and Accessories
- 1x Fluke 5615-12-P Secondary Reference Temperature Probe
- 1x Fluke 2373-LTC Lemo to Universal
- 1x Carry Case
Fluke 1523-256 1-Channel Handheld Thermometer Kit 4 Technical Specifications
Please find the available technical specifications for the instruments included in this kit.
Fluke 1523-256 1-Channel Handheld Thermometer Technical Specifications
Please note that the Fluke 1523-256 1-Channel Handheld Thermometer has a sister product, the Fluke 1524-256 2-Channel Handheld Thermometer, please find below the technical specifications for both of these models.
Specification | 1523 | 1524 |
Input channels | 1 | 2 |
Resolution | PRTs and thermistors: 0.001° thermocouples: 0.01° | |
Logging | 25 readings with statistics | 25 readings with statistics15,000 time and date stamped |
Sample interval (normal) | 1 second | 1 second (simultaneous measurement) |
Typical sample interval (fast mode) | 0.3 seconds | |
Sensor types | PRTs, RTDs, thermistors, and thermocouples | |
Thermocouple types | C, E, J, K, L, M, N, T, U, B, R, S | |
Statistics | Maximum, minimum, average, standard deviation | |
Trending | Scale: ±10 °C (18 °F), ±1 °C (1.8 °F), ± 0.1°C (0.18 °F), ± 0.01°C (0.018 °F), 10 minutes of real-time data | |
Power requirements | 3 AA alkaline batteries, 12 V dc universal power supply | |
Size (HxWxD | 96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in) | |
Weight | 0.65 kg (1.4 lb) | |
Computer interface | RS-232, 9940 I/O ToolKit software included | |
Safety | EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 | |
Environmental conditions for best accuracy: 13 °C to 33 °C (55.4 °F to 91.4 °F) | ||
Millivolt range and accuracy | –10 mV to 75 mV, ± (0.005 % + 5 µV) | |
Internal reference junction compensation | ± 0.2 °C (± 0.36 °F) | |
Resistance range and accuracy |
0 Ω to 400 Ω ± (0.004 % + 0.002 Ω) 200 Ω to 50 kΩ ± (0.01 % + 0.5 Ω) 50 kΩ to 500 kΩ ± (0.03 %) |
|
Temperature coefficient, voltage: –10 °C to 13 °C , +33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) |
± (0.001 %/°C + 1 mV/°C) | |
Temperature coefficient, resistance: –10 °C to 13 °C ,+33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) |
0.0008 %/°C + 0.0004 Ω (0 Ω to 400 Ω) 0.002 %/°C + 0.1 Ω (0 Ω to 50 kΩ) 0.06 %/°C + 0.1 Ω (50 kΩ to 500 kΩ) |
|
Excitation current, resistance | 1 mA (0 Ω to 400 Ω) 10 µA (0 Ω to 50 kΩ) 2 µA (50 kΩ to 500 kΩ) | |
*See technical manual for sample interval details by sensor type and number of inputs. |
Fluke 5615-12-P Secondary Reference Temperature Probe Technical Specifications
Temperature range | –200 °C to 420 °C | |
Nominal resistance at 0 °C | 100 Ω ± 0.10 Ω | |
Temperature coefficient | 0.0039250 Ω/Ω/°C | |
Accuracy[1] | ± 0.024 °C at –200 °C | |
± 0.012 °C at 0 °C | ||
± 0.035 °C at 420 °CC | ||
Short-term repeatability[2] | ± 0.009 °C at 0.010 °C | |
Drift[3] | ± 0.007 °C at 0.010 °C | |
Sensor length | 28 mm (1.1 in) | |
Sensor location | 6.9 mm ± 3.3 mm from tip (0.27 in ± 0.13 in) | |
Sheath dimensions | 305 mm x 6.35 mm (12.0 in x 0.250 in) | |
Sheath diameter tolerance | ± 0.127 mm (± 0.005 in) | |
Sheath material | Inconel™ 600 | |
Minimum insulation resistance | 1000 MΩ at 23 °C | |
Transition junction temperature range[4] | –50 °C to 200 °C | |
Transition junction dimensions | 71 mm x 13 mm dia (2.8 in x 0.5 in) | |
Maximum immersion length | 254 mm (10 in) | |
Response time[5] | 9 seconds typical | |
Self heating (in 0 °C bath) | 50 mW/°C | |
Lead-wire cable type | PTFE insulated with PTFE jacket, 22 AWG | |
Lead-wire length | 183 cm (72 in or 6 ft) | |
Lead-wire temperature range | –50 °C to 200 °C | |
Calibration | NVLAP-accredited calibration included, lab code 200706-0. Please see calibration uncertainty table and its explanation of changeable uncertainties. | |
Notes | [1] Includes calibration and 100 hr drift (k=2) | |
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2) | ||
[3] After 100 hrs at max temp, 95 % confidence (k=2) | ||
[4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch. | ||
[5] Per ASTM E 644 | ||
NVLAP† Calibration Uncertainty |
||
Temperature | Expanded Uncertainty (k=2) | |
–196 °C | 0.024 °C | |
–38 °C | 0.011 °C | |
0 °C | 0.010 °C | |
200 °C | 0.018 °C | |
420 °C | 0.029 °C | |
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 |