The FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger features a T-type temperature probe with a 1m cable and can be used to monitor the temperature of vaccines stored in cryogenic dry ice. It has a temperature range of -100°C to +100°C and will wirelessly stream data to the EasyLog Cloud or a PC with the EasyWiFi Software installed.
There are two versions of the FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger available: a single-channel model and a dual-channel model. Please see the table below for more information.
SKU | Title | Channels |
EL-WiFi-ULT | FilesThruTheAir Cryogenic Vaccine Temp Data Logger - 1 Channel | Single |
EL-WiFi-DULT | FilesThruTheAir Cryogenic Vaccine Temp Data Logger - 2 Channel | Dual |
If the FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger loses communication with the Wi-Fi network it will continue to log results until it is able to reconnect. In this situation, it can log readings for a maximum of 30 days at a logging rate of 10 seconds.
Users can configure the FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger to log readings every 10s, 10mins or 12hrs. They can also set high and low alarms; when these alarms are triggered the FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger will issue an alert.
The FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger can be configured using a PC and the EasyLog WiFi Software or the EasyLog Cloud App. Additionally, this software and app can be used to view, graph, and analyse logged data.
Finally, this free-standing, IP40-rated temperature data logger can be attached to a wall or surface using the supplied bracket. The FilesThruTheAir EL-WiFi-D/ULT Cryogenic Vaccine Temperature Data Logger is simple-to-use and reliable; it has been specifically designed for monitoring vaccines that must be stored in dry ice.
For further information regarding FilesThruTheAir's Wireless Data Loggers and the EasyLog Cloud, please see the video below.
Minimum | Typical | Maximum | Unit | |
Battery life | >6 | Months | ||
Temperature measurement range | -100 (-148) | +100 (+212) | °C (°F) | |
USB supply voltage (@500mA) | 4.5 | 5.0 | 5.5 | Vdc |
Operating temperature range (logger) | -20 (-4) | +60 (+140) | °C (°F) | |
Logging period (user configurable) | 10 sec | 10 min | 12 hrs | |
Transmission period (user configurable) |
1 min | 1 hr | 24 hrs | |
Temperature measurement resolution | 0.1 (0.2) | °C (°F) | ||
Temperature display resolution | 0.1 | |||
Accuracy | ±2.5 (±4.5) | °C (°F) | ||
IP Rating | IP40 | |||
(Bung fully inserted, not permanently powered, thermocouple connector fitted and fully inserted, device mounted vertically.) |
||||
Dimensions | 82 x 70 x 36mm (3.22 x 2.75 x 1.41”)** | |||
Probe dimensions | 300 x Φ 1.5mm, 1m cable (11.8 x Φ 0.05, 39.3” cable) |