Connect up to two thermocouples to this logger to constantly monitor temperature from a fixed point.
Compatible with K, J and T type thermocouples, the EL-GFX-DTC accurately measures temperature, logs it in memory and can also display values on its built-in LCD screen. This screen can be used to interact with the logger in the field and can show real-time analysis of data as a summary (showing highest and lowest readings/alarm conditions) or as a graph that updates automatically when new data is added.
Done logging? Simply plug the EL-GFX-DTC into a PC and download into the included software for further analysis. This software can also be used to configure the logger’s alarm thresholds, logging rate and more.
This logger is powered by AA batteries which will last approximately for four months before needing replacing.
Specifications | Minimum | Typical | Maximum | |
Probe Measurement Range | K Type | -200°C (-328°F) | +1350°C (+2462°F) | |
J Type | -200°C (-328°F) | +1190°C (+2174°F) | ||
T Type | -200°C (-328°F) | +390°C (+734°F) | ||
Internal Resolution | - | 0.1°C (0.1°F) | - | |
Accuracy (Overall Error) | - | ±0.5°C (±0.9°F) | - | |
Logging Rate | Every 2 seconds | - | Every 1 hour | |
Operating Temperature Range | -10°C (-14°F) | - | +40°C (+104°F) | |
Battery Life | - | 4 months (varies depending on functions used) | - |