Suitable for measuring diesel vehicle transportation/service depots, fossil fuel burning, traffic emission, charcoal heating, manufacturing and more, the Extech VPC300 is a an advanced particle counter which can simultaneously measure and display 6 particle size channels down as low as 0.3μm.
Fully configurable to the user's specifications, the VPC300 allows adjustments to sampling time, counting data and even has a programmable delay function. MAX/MIN, DIF and AVG recording functions are also available, as is a date/time setup system. This meter is also suitable for continuous measurements and comes supplied with a tripod for hands-free mounting.
An on-board digital camera/video system on the VPC300 can be used to take reference videos and images which are stored directly in internal memory. Up to 5000 individual testing records can be directly stored on the tester and a 20 minute video can also be stored alongside them. Afterwards, Extech's included software can be used to generate reports on PC with videos, photos and data points.
This meter is calibrated to NIST-traceable standards by Extech at the time of manufacturing.
Extech VPC300 Particle Counter Key Features
- Choice of particle sizes - 0.3, 0.5, 1.0, 2.5, 5.0 and 10μm
- Cumulative, differential and concentration counting modes
- Simultaneous measure particle sizes from 6 points
- Calculation of air temperature, humidity, dew point and wet bulb
- 2.8" TFT full colour LCD display
- Built-in 320 x 240 pixel camera capable of capturing videos and still images
- Store up to 5000 testing records in internal memory
- Selectable sampling time and programmable delay
- MAX/MIN/AVG/DIF recording
- Automatic power off when left idle
- Language selection: choose between English, French, German and Spanish
- Tripod mount included for continuous, hands-free monitoring
- Mini USB connection with PC
- Create reports using the meter and Extech's included PC software
- MicroSD card slot (max 8GB) to expand memory (SD card not included)
What's Included?
- Extech VPC300 Particle Counter
- NIST-Traceable Calibration Certificate (Manufacturer-Issued)
- Universal AC Adapter/Charger w/ Multiple Plugs
- USB Cable
- PC Software
- Tripod
- Filter
- 7.4V NiMH Rechargeable Battery
- Hard Case
Extech VPC300 Technical Specifications
Particle Counter | |
Channels | Six: 0.3, 0.5, 1.0, 2.5, 5.0, 10μm |
Flow Rate | 0.1ft³/min (2.83L/min) |
Counting Efficiency | 50% at 0.3μm; 100% for particles > 0.45μm |
Coincidence Loss | 5% at 2,000,000 particle per ft³ |
Particle Count Modes | Cumulative, differential and concentration |
Air Temp/Relative Humidity | |
Air Temperature Range | 32 to 122°F (0 to 50°C) |
Dew Point/Wet Bulb Temperature Range | 32 to 122°F (0 to 50°C) |
Relative Humidity Range | 0 to 100% RH |
Air Temperature Accuracy | ±0.9°F (0.5°C) 50 to 104°F (10 to 40°C) ±1.8°F (1.0°C) all other ranges |
Dew Point/Wet Bulb Temperature Accuracy | ±0.9°F (0.5°C) 50 to 104°F (10 to 40°C) ±1.8°F (1.0°C) all other ranges |
Relative Humidity Accuracy | ±3% RH from 40 to 60% RH ±3.5% RH from 20 to 40% RH and from 60 to 80% RH ±5.0% RH from 0 to 20% RH and from 80 to 100% RH |
General Specifications | |
Display | 2.8" 320 x 240 pixel TFT colour LCD with backlight |
Low Battery Indication | Battery symbol appears on the LCD |
Micro-SD Card Slot | Accommodates up to an 8GB SD card |
Meter Internal Memory | 5000 sample records; 74MB |
Power Supply | Rechargeable 7.4V Lithium-Ion polymer battery |
Battery Life | Approx 4 hours of continuous use |
Battery Charge Time | 2 hours approx with AC adapter/charger |
Operating Temperature | 32 to 122°F (0 to 50°C) |
Operating/Storage Humidity | 10 to 90% RH non-condensing |
Storage Temperature | 14 to 140°F (-10 to 60°C) |
Dimensions | 9.4 x 3.0 x 2.2" (240 x 75 x 57mm) |
Weight | 1.26lbs (570g) |
Reviews
SKU | VPC300 |
---|---|
Manufacturer Part Code | VPC300 |
Model Ref | Extech VPC300 |
Country of Origin | China |