This Chauvin Arnoux tachometer can be used to accurately gain measurements of RPM in various different applications.
A handheld, portable device, the Chauvin Arnoux CA1725 is designed for use in just one hand and features a rotary dial for easily adjusting the modes of the tester. The device also includes a selection of button for adjusting commonly used features and the main body of the unit includes a large LCD screen that can be easily viewed when testing procedures are taking place.
This tachometer is both a non-contact device and can also be used to carry out contact tachometer measurements via the use of various fitting types (see tech specs for more information on these). The non-contact mode of this tachometer requires the use of a reflective strip and an infrared beam from the tester. The reading is calculated based on the infrared light bouncing from the rotating piece of machinery.
Chauvin Arnoux CA1725 Tachometer Key Features
- RP measurements in tr/min and m/min
- Frequency and ms measurements
- Duty %, RPM and ft/min
- Portable device
- Infrared and contact RPM measurements
- MIN/MAX
- Smoothing function
- Data hold function
Full List of Products Included
- Chauvin Arnoux CA1727 Tachometer
- Hard Carrying Case
- 1 x FRB F Connector
- 1 x 9V Battery
- 1 Set of 15 Reflective Strips (0.1m Length)
- 1 x Operating Manual
- 1 x Quick-Start Up Guide
Chauvin Arnoux CA1725 Technical Specifications
RPM Function | Range: 60 to 100,000 rpm Resolution: 0.006 to 6 according calibre Accuracy: 1 x 10-4 of reading ± 6 counts |
m/min Function | Range: 6 to 10,000 m/min Resolution: 0.006 to 6 depending on calibre Accuracy:1 x 10-4 of reading ± 1 increment |
Hz Function | Range: 1 to 10,000 Hz Resolution: 0.004 to 0.4 depending on calibre Accuracy: 4 x 10-5 of reading ± 4 counts |
ms Function | Range: 0.1 to 1,000 ms Resolution: 0.0003 to 0.03 depending on calibre Accuracy: 1 x 10-4 of reading ± 5 counts |
Duty Cycle Function | Range: 0.1 to 100 % Resolution: 0.1 to 1 % depending on calibre Accuracy: 0.1 % of scale from 0.2 Hz to 50 Hz • 0.2 % of scale from 50 Hz to 125 Hz • 1 % of scale otherwise |
Count Function (CA 1727) | Range: 0 to 100,000 events Accuracy: ± 1 event |
Power Supply | 9V |
Battery Life | 250 x 5 min measurements with optical sensor 600 x 5 min measurements with external sensor |
Memory | 4,000 points |
Dimensions | 21 x 72 x 47 mm |
Weight | 250g |
Leakproofing | IP 51 |
Environment | Storage: -20 °C to +70 °C 95 % RH Operation: 0 to 55°C 90 % RH |
Optical Sensor Specs | Reflective area: 10 to 90 % of the target area. Measurement distance: 1 to 50 cm. The maximum distance is given for a reflective adhesive strip with a minimum area of 10 cm2. Measurement angle: ± 15° in relation to the perpendicular of the reflective surface. |
Mechanical Adapter Specs | End-fittings: elastomer with a hardness of 80 shores Pressure on moving part: between 2 and 40 N. Maximum speed: 10,000 rpm. Service life: approx. 1,000 hours at 3000 rpm with a pressure of 20 |
Conical End Fitting Accessory Specs | Minimum diameter of measurement shaft: 5 mm. |
Cylindrical End-Fitting Accessory Specs | Speed measurements on shafts with a diameter greater than 5 mm or flat-ended shafts |
End-Fitting Accessory (with Wheel) Specs | Wheel diameter: 30.183 mm. Wheel development: 10 cm ± 0.1 mm |
Reviews
SKU | P01174810 |
---|---|
Manufacturer Part Code | P01174810 |
Country of Origin | France |