All Chauvin Arnoux Products

Chauvin Arnoux CA 1510 Indoor Air Quality Tester

SKU
CA1510
Rating:
0%
(0)
Reward Points
:
414
£414.00 £345.00
1 Option in Stock
Free UK Delivery*
Advice & Support Call 01642 931329 or contact us
Price Match We won't be beaten, let us know
Chauvin Arnoux CA 1510 Indoor Air Quality Tester

The Chauvin Arnoux CA 1510 Indoor Air Quality Tester can be used to monitor CO2 concentration, air humidity and air temperature. It provides an initial indication of the air quality and can be used alongside pollution analysis to ascertain the effectiveness of air renewal systems.

The Chauvin Arnoux CA 1510 Indoor Air Quality Tester can measure CO2 concentrations between 0 and 5000ppm to an accuracy of ±50ppm or 3% of the measurement. It can also measure air temperature between -10°C and +60°C to an accuracy of ±0.5°C and air humidity between 5% to 95%RH to an accuracy of ±2%RH. To maintain measurement accuracy, the Chauvin Arnoux CA 1510 Indoor Air Quality Tester is supplied with an adapter, tube and software for on-site calibration. The user can adjust the zero and the 1700ppm threshold as required.

Up to three parameters may be simultaneously displayed on the Chauvin Arnoux CA 1510 Indoor Air Quality Tester’s two-colour, backlit screen. The Chauvin Arnoux CA 1510 Indoor Air Quality Tester also includes integrated technology which interprets parameter readings to indicate the level of comfort within the room. It features two interpretation modes:

  • 1D Mode: indicates CO2 confinement; if the CO2 concentration is between 1000ppm and 1700ppm, the Chauvin Arnoux CA 1510 Indoor Air Quality Tester will alert the user using visual (display colour and pictogram) and audible alerts
  • 3D mode: the Chauvin Arnoux CA 1510 Indoor Air Quality Tester will assess the comfort level based on the CO2 concentration and hygrothermal data

The Chauvin Arnoux CA 1510 Indoor Air Quality Tester is capable of logging up to one-million readings which are saved as part of several measurement campaigns. The Chauvin Arnoux CA 1510 Indoor Air Quality Tester’s display and keyboard may be locked during recording. Data can be transferred via a USB interface or wirelessly via Bluetooth for analysis using the supplied software or Chauvin Arnoux’s 1510 Android App. The supplied software can be used to view readings remotely, display data in a graph or table, export measurements to Excel, and generate reports.

Finally, this portable air quality monitor features a simple user interface and supports various mounting possibilities: the Chauvin Arnoux CA 1510 Indoor Air Quality Tester is magnetic, alternatively, it can be suspended or mounted using wall support, wall support with a padlock insert, or a desktop stand.

The Chauvin Arnoux CA 1510 Indoor Air Quality Tester is an easy-to-use and ergonomic air quality monitor available in black or white.

Chauvin Arnoux CA 1510 Indoor Air Quality Tester Key Features

  • Monitor CO2 concentration, air humidity and air temperature
  • Provides an initial indication of air quality
  • Use alongside pollution analysis to ascertain the effectiveness of air renewal systems
  • CO2 range: 0 to 5000ppm
  • CO2 accuracy: ±50ppm or 3% of the measurement
  • Air temperature: -10°C to +60°C
  • Air temperature accuracy: ±0.5°C
  • Air humidity: 5% to 95% RH
  • Air humidity accuracy: ±2%RH
  • Adapter, tube and software for on-site calibration: adjust the zero and the 1700ppm threshold
  • Two-colour, backlit display
  • View up to three parameters on the display
  • Integrated comfort level technology: interprets parameters to indicate the level of comfort within the room
  • 1D mode: ascertains the comfort level based on CO2 levels; issues visual and audible alerts when the level of CO2 is between 1000ppm and 1700ppm
  • 3D mode: determines the comfort level based on CO2 concentration and hygrothermal data
  • Log up to 1 million readings
  • Lock the display and keyboard during recording
  • Transfer data via a USB interface or wirelessly via Bluetooth
  • Analyse and export data and create reports using the supplied software
  • Compatible with Chauvin Arnoux’s 1510 Android App (NOT included)
  • Portable
  • Simple user interface
  • Magnetic
  • Can be suspended
  • Can be fixed into position using wall support, wall support with a padlock, or a desktop stand
  • Easy-to-use and ergonomic air quality monitor available in black or white

What’s Included?

  • Chauvin Arnoux CA 1510 Indoor Air Quality Tester (Your Choice of Black or White)
  • 2x LR6 Batteries
  • USB Mains Adapter
  • 8m USB-micro USB Cable
  • Quick Start Guide
  • Mini-CD Containing the Data Logger Transfer Software and User Manual
  • Verification Attestation

Chauvin Arnoux CA 1510 Indoor Air Quality Tester Technical Specifications 

CO2 Measurement
Measuring principle Non-diffusive infrared (NDIR) technology
Type of sensor Double-beam infrared cell sensor
Measurement range 0 to 5,000 ppm
Uncertainty ± 50 ppm ± 3 % of value measured
Response time, 63 % < 200 seconds
Resolution 1 ppm
Temperature measurement
Type of sensor CMOS
Measurement range -10 °C to +60 °C
Accuracy ± 0.5 °C
Resolution 0.1 °C
Humidity Measurement 
Type of sensor Capacitive
Measurement range 5 to 95 % RH
Accuracy ± 2 % RH
Resolution 0.1 % RH
Types of Use 
Point Measurement Monitoring Quick measurement and display of the CO2, temperature and relative humidity values
Monitoring 1D mode: indication of CO2 confinement
Visual (two-colour backlighting & pictograms) and/or audible indication of high
confinement when the CO2 concentration is between 1,000 ppm and a 1,700 ppm
threshold.
3D mode: indication of optimum comfort zone on the basis of the hygrothermal criteria
and the CO2 concentration.
Energy-saving (ECO): for fixed use on battery power, the product performs
measurements every 10 minutes over a programmable time range for a battery life of
up to one year.
Logger Triggering of programmed recording (P_REC): The start date, recording rate and end date can be customized with the PC software
or the Android application. Possibility of locking the display in this mode (no values
displayed).
  Manual triggering (M_REC): Manual start and stop controls on the product.
Recording is performed at the rate of the mode currently selected.
Specifications 
Recording interval Customizable from 1 minute to 2 hours
Storage More than 1 million measurements
Buzzer Yes
Units °C or °F
Backlighting
Display Hold function
Min-Max
Yes
Automatic power-off Yes (in portable mode only)
General Specifications 
Product operating range Temperature: -10 °C to +60 ° C – Humidity: 5 to 95 %RH
Dimensions / weight 125 x 65.5 x 32 mm / 190 g with batteries
Protection IP40
Compliance IEC 61010-1 for 50 V in Category II – IEC 61326-1 – Compliance with French decree no. 2012-14
Power supply - Alkaline batteries: 2 x 1.5 V AA / LR6 or rechargeable battery
- Connection to mains possible with mains / micro USB adapter supplied as standard
Interfaces 2 communication modes possible:
- Bluetooth wireless communication
- USB link; the product is then recognized as a USB key for easy file transfer
Mounting C.A 1510 casing equipped with a magnet, a wall-suspension system and a slot for hanging the product. A wall support for use with a padlock
(padlock not supplied) is available as an accessory, as is a desktop stand (supplied as standard with the C.A 1510W).
Functions of the Data Logger Transfer
software supplied as standard

– Graphic representation or as table of values

– Data export

– Real-time mode

– Report generation

Chauvin Arnoux CA 1510 Indoor Air Quality Tester - Datasheet Datasheets Chauvin Arnoux CA 1510 Indoor Air Quality Tester - Quick Start Guide Manuals & Guides Chauvin Arnoux CA 1510 Indoor Air Quality Tester - User Manual Manuals & Guides Chauvin Arnoux CA 1510 IAQ Tester - Air Quality & Risk of Contamination Case Study Case Studies & Applications Chauvin Arnoux CA 1510 IAQ Tester - COVID-19 Case Study Case Studies & Applications Chauvin Arnoux CA 1510 Indoor Air Quality Tester - Inspections & Repairs Notes & Other

Reviews

Additional Information
SKU CA1510