Beha-Amprobe's 37XR-A digital multimeter includes all of the standard electrical measurements you'll need on a daily basis, plus dedicated component and logic tests allowing you to easily troubleshoot electronics either on the bench, or when you're out working.
A reliable, high accuracy digital multimeter with true RMS included as standard, the 37XR-A will maintain its accuracy even when you're testing on irregular AC circuits. The meter includes a 10,000 count digital LCD display and a simple rotary dial for changing between different testing procedures.
The 37XR-A can measure DC voltage to 1000V, AC voltage to 750V, AC/DC current to 10A, resistance to 40MΩ, capacitance to 400μF, inductance to 40H, frequency to 10MHz, duty cycle from 0 to 90% and also tests continuity, diode and logic (TTL and CMOS).
This multimeter also includes several other useful functions including the ability to record high/low values taken over a period of time with the MIN/MAX function, the ability to determine the most common reading with the average function and dedicated peak and data hold functions for storing values temporarily on the screen.
The Beha-Amprobe 37XR-A is CAT III 600V and CAT II 1000V rated and includes a 3 year warranty as standard.
Beha-Amprobe 37XR-A True RMS Digital Multimeter Key Features
- Reliable digital multimeter ideal for electrical and electronic testing, including testing of components
- True RMS - ensures accuracy is maintained even when testing on unusual AC circuits
- Input warning tone when test leads are inserted incorrectly
- 10,000 count digital display
- Measure AC voltage to 1000V, DC voltage to 750V
- Resistance measurements to 40MΩ
- Continuity test with integrated beeper
- Inductance measurements to 40H
- Frequency measurements to 40MHz
- Capacitance measurements to 400μF
- AC/DC current measurements to 10A
- Logic test (TTL and CMOS)
- Diode test
- MIN/MAX mode calculates the maximum and minimum values taken over a period of time
- Average function
- Relative and peak functions
- Data hold function temporarily stores current value shown on the screen
- Automatic power off helps save battery life
- Separate door for easy access to battery and fuse
- Magne-grip holster included with magnetic hanging strap
- CAT III 600V and CAT II 1000V rated
- Three year warranty
What's Included?
- Beha-Amprobe 37XR-A Multimeter
- Test Leads + Threaded Alligator Clips (TL36A)
- 9V Battery (Installed)
- Magne-Grip Holster with Magnetic Hanging Strap (H2-XRA)
- Spare Fuse
- Operator's Manual
Optional Accessories
- VC221B - Padded VInyl Carrying Case
- DL248D - Deluxe Test Lead Kit
- DC205C - Heavy Duty Case
- HV231-10A - High Voltage Probe
- TP254A - High Temperature Probe
- PASS Oxford Black Small Test Equipment Carry Case - PASSCASE-SM, typically £14.99
Beha-Amprobe 37XR-A Technical Specifications
Full list of technical specifications can be found in the attached user manual.
General Specifications | |
---|---|
Display | 4 and 3/4 digit liquid display (LCD) with a 41 segment analogue bargraph |
Autoranging | 9999 counts |
Manual Ranging | 9999 counts |
Polarity | Automatic, positive implied, negative polarity indication |
Overrange | (OL) or (-OL) displayed |
Zero | Automatic |
Low Battery Indication | Symbol displayed when the battery voltage drops below the operating level |
Auto Power Off | Approx 30 minutes |
Measurement Rate | 2 times per second, nominal |
Operating Environment | 0 to 50°C at <70% RH |
Storage Temperature | -20 to 60°C, 0 to 80% RH with battery removed from meter |
Temperature Coefficient | 0.1 x (specified accuracy) per °C. (0 to 18°C, 28 to 50°C) |
Altitude | 2000m (6562 feet) |
Power | Single standard 9V battery, NEDA 1604, JIS 006P, IEC 6F22 |
Battery Life | 75 hours typical with carbon-zinc. 150 hours typical with alkaline. Using the backlight will decrease battery life |
Dimensions (H x W x D) | 196 x 92 x 60mm |
Weight | 482g (with battery and holster) |
Electrical Specifications | |
DC Voltage | |
Ranges | 1000mV, 10V, 100V, 1000V (auto/manual ranging) |
Resolution | 100μV |
Accuracy | ±(0.1% rdg + 5 digits) |
Input Impedance | 10MΩ |
Overload Protection | 1000V DC or 750V AC rms |
AC Voltage (True RMS) | |
Ranges | 1000mV, 10V, 100V, 750V (auto/manual ranging) |
Resolution | 100μV, minimum reading on 1000mV range: 14mV |
Accuracy | ±(1.2 % rdg + 10 dgts) 45 Hz to 500 Hz ±(2.0% rdg + 10 dgts) 500 Hz to 2 kHz ±(2.0% rdg + 10 dgts) 45 Hz to 1 kHz on 750 V range |
Peak Hold Accuracy | ±(3.0% + 200 digits) on 100V, 750V range 1000mV, 10V ranges unspecified |
Crest Factor | ≤3 |
Input Impedance | 10MΩ AC coupled true rms specified from 5 to 100% of range |
Overload Protection | 1000V DC or 750V AC rms |
DC Current | |
Ranges | 100μA, 1000μA, 10mA, 100mA, 400mA, 10A (auto/manual ranging) |
Resolution | 0.01μA |
Accuracy | ±(0.5 % rdg +10 dgts) on 100μA range ±(0.5 % rdg + 5 dgt) on 1000μA to 400mA ranges ±(1.5 % rdg + 10 dgts) on 10A range |
Input Protection | 0.5A/1000V fast blow ceramic fuse 6.3×32 mm on μA/mA input 10A/1000V fast blow ceramic fuse 10×38 mm on 10A input 10A input: 10 A for 4 minutes maximum followed by a 12 minute cooling period |
Burden Voltage | μA range: 1mV/1μA mA range: 10mV/1mA A range: 35mV/1A |
AC Current (True RMS, 45 to 1kHz) | |
Ranges | 100μA, 1000μA, 10mA, 100mA, 400mA, 10A (Auto/Manual ranging ) |
Resolution | 0.01μA |
Accuracy | ±(1.5% rdg +10 dgts) on 100μA to 100mA ranges ±(2.0% rdg +10 dgts) on 400mA range ±(2.5% rdg + 20 dgts) on 10A range Peak Hold accuracy: ±(3.0 % + 200 dgts) 100μA range unspecified |
Input Protection | 0.5A/1000V fast blow ceramic fuse 6.3×32 mm on μA/mA input 10A/1000V fast blow ceramic fuse 10×38 mm on 10A input 10A input: 10 A for 4 minutes maximum followed by a 12 minute cooling period |
Crest Factor | 3≤ |
Burden Voltage | See DC current |
Resistance | |
Ranges | 1000Ω, 10kΩ, 100kΩ, 1000kΩ, 10MΩ, 40MΩ (auto/manual ranging) |
Resolution | 100mΩ |
Accuracy | ±(0.5% rdg + 8 dgts) on 1000Ω to 1000kΩ ranges ±(1.0% rdg + 10 dgts) on 10MΩ range ±(2.0% RDG + 10 dgts) on 40MΩ range |
Open Circuit Voltage | -0.45V DC typical |
Overload Protection | 1000V DC or 750V AC rms |
Capacitance | |
Ranges | 40nF, 400nF, 4μF, 40μF, 400μF (3999 counts) (auto/manual ranging) |
Resolution | 0.01nF |
Accuracy | ±(3.0% rdg + 10 dgts) on 40nF, 400μF ranges ±(3.0% rdg + 5 dgts) on 400nF to 40μF ranges |
Test Voltage | <1V |
Test Frequency | 1.3Hz on 40nF to 40μF ranges; 0.7Hz on 400μF range |
Input Protection | 0.5A/1000V fast blow ceramic fuse 6.3 x 32mm on μA/mA input |
Reviews
SKU | 2727808 |
---|---|
Manufacturer Part Code | 2727808 |
Model Ref | Amprobe 37XR-A |
Country of Origin | Taiwan |